45
Select “Power” to check the IEPE power supply of your Fieldpaq
Ⅱ
. To
conduct the checking, discount all sensors and select Verify. In normal
condition, the power supply of all channels should be around 24.7 Volts. If the
result shows abnormal voltage levels, please contact Benstone Instruments
for assistant.
TTL Out :
The TTL Out allows you to have the Fieldpaq
Ⅱ
send a TTL signal out via the
auxiliary channel to another device such as a strobe. This feature is based on
the cursor position in the frequency domain function such as spectrum, Auto
Power Spectrum, or FRF. This feature requires an optional TTL out cable. An
example of application is when you have two rotating devices at similar
frequencies, and by strobing each device with a precise frequency, you can
determine which frequency belongs to which rotating device.
RPM ON :
Selecting the RPM On, will display the RPM value of the external trigger
channel in the status bar.