4.3 Measuring ranges
Device description
5225 / 04/2022 en
BENNING IT 100
22
Measuring accuracy
The measuring accuracy is specified as the sum of the following:
• Relative part of the measured value
• Number of digits (counting steps of the last digit)
The stated measuring accuracy is specified for 1 % to 100 % of the final measuring range value
and applies at a temperature of 23 °C ±5 °C and a relative air humidity lower than 80 %. In case
of deviating temperatures, observe the temperature coefficient by adding the following value to
the specified measuring accuracy:
0.2 [1/°C] x specified measuring accuracy x difference to reference temperature range [°C]
4.3.1
Voltage ranges
AC voltage ranges (V‑AC)
Overload protection: 600 V‑AC / V‑DC
Measuring range
Resolution
Measuring accuracy
60 V
0.01 V
±(2 % + 5 digits)
600 V
0.1 V
Table 5:
AC voltage ranges (V-AC)
• Frequency range: 45 … 500 Hz
DC voltage ranges (V‑DC)
Overload protection: 600 V‑AC / V‑DC
Measuring range
Resolution
Measuring accuracy
60 V
0.01 V
±(2 % + 5 digits)
600 V
0.1 V
Table 6:
DC voltage ranges (V-DC)
4.3.2
Resistance ranges
Measuring ranges for resistance measurement
Overload protection: 600 V‑AC / V‑DC
Measuring range
Resolution
Measuring accuracy
600 Ω
0.1 Ω
±(2 % + 5 digits)
6 kΩ
0.001 kΩ
60 kΩ
0.01 kΩ
600 kΩ
0.1 kΩ
Table 7:
Measuring ranges for resistance measurement (Ω)