6.7 Replacing the fuse
Technical data
5225 / 04/2022 en
BENNING IT 100
38
7
Technical data
Protection class
II (double or reinforced insulation)
Contamination level
2
Protection category
(DIN VDE 0470-1, IEC / EN 60529)
IP 30
1st digit: 3 = protection against access to
dangerous parts and protection against solid
impurities (diameter >2.5 mm)
2nd digit: 0 = no protection against water
Overvoltage category
• CAT IV 600 V to earth
Housing dimensions
(length x width x height)
206 mm x 90 mm x 51 mm
Weight (batteries and protective rubber holster
included)
0.51 kg
Battery life (alkaline batteries)
• approx. 2 600 low-resistance
measurements acc. to IEC /
DIN EN 61557‑4 (1 Ω, with a measuring
duration of 5 s)
or
• approx. 1 200 insulating resistance
measurements acc. to IEC /
DIN EN 61557‑2 (1 MΩ, at 1 000 V and
with a measuring duration of 5 s)
Set of safety measuring lines (item no.: 10230645)
Standard
IEC / DIN EN 61010‑031 (VDE 0411‑031)
Overvoltage category
(only applies to the safety measuring lines,
additionally observe the limitations of the
device)
• With attachable protective cap or alligator
clip:
– CAT III 1 000 V to earth
– CAT IV 600 V to earth
• Without attachable protective cap:
– CAT II 1 000 V to earth
Protection class
II (double or reinforced insulation)
Contamination level
2
Max. rated current
10 A
Length
1.3 m
Operation
Max. barometric altitude
2 000 m
Operating temperature
0 … 40 °C (do not permanently expose the
device to sunlight)
Max. relative air humidity
80 % RH (0 … 40 °C), non-condensing
Operating conditions
To be used inside buildings in dry
environments
Storage
(remove the batteries from the device)
Ambient temperature
-10 … 60 °C (do not permanently expose the
device to sunlight)
Max. relative air humidity
80 % RH
Table 11:
Technical data