BENNING MA 4 measuring adapter
12
6.2.2. R
Insu
1
–
(LN-PE)
Before testing
Disconnect the test sample from the mains supply.
Switch the rotary switch (7) of the
measuring
adapter
to position "
L1-L2-L3
".
The rotary switch (8) of the
measuring adapter
must be set to the position for
appliance testing
.
Test procedure (VDE 0701-0702)
Portable test sample
Connect the CEE connector of the test sample to the
CEE socket of the
measuring adapter
.
Start the R
Insu
measurement on your appliance tester.
As a result, you will obtain the measured R
Insu
value of the test
sample.
Application
Portable test sample
Connection diagram
Circuit diagram
6.2.3. R
Insu
2
–
(sec.-PE)
Portable test sample
Connection diagram
Circuit diagram
6.2.4. R
Insu
3
–
(LN-sec.)
Portable test sample
Connection diagram
Circuit diagram