• Wipe off the first drop
of blood.
• Massage your finger
gently till a round drop
of blood forms.
9
• Touch the blood sample with
strip inlet.
• The sample will be drawn
automatically.
• After count down, the result will display on the screen.
• Then eject the used strip into a biohazard waste container.
• The meter will turn off automatically.
11
10
Note:
Make sure you have enough sample for sufficient test.
If the blood smears, please wipe off and squeeze another drop of blood.
Note:
Insufficient fill up of blood sample could lead to inaccurate or failed test result.
Note:
• Please code your meter when you first use or open a new vial of strips.
If you have any question, please read Chapter 2.3.
• Please follow the instruction while testing. Improper operation may result in inaccurate
test result or damage the system.
• Please finish the test within 5 minutes or the meter will turn off automatically.
• The meter will not turn on if you insert the wrong end or wrong side of strip.
• Do not use expired strip.
Warning :
●
Lancets cannot be reused.
●
Please always use certified lancets to ensure safety.
3.3 Alternate Site Testing (AST)
The meter allows you to test glucose from your fingertip, palm or forearm. Blood obtained
from palm and forearm may reduce pain during sampling process, but the glucose level
changes faster in palm and forearm. These differences may cause wrong medical decision.
Note:
Please consult healthcare professional before AST sampling.
Suitable timing to acquire blood sample from alternate sites:
• Routinely before meal
• Prior to or two hours after a meal/a short-acting or rapid-acting insulin analogue/exercise.
11
12
Fill up ok
Fill up FAILED
Glucose
∕
Uric Acid
∕
Total Cholesterol
0.7
0.8
GLU
UA
CHOL