Product description
12
CP9xx-Kit_D00399_00_M_XXEN/07.2019
3.5 Function
CP9… alarm indicator and operator panels are integrated into the existing EDP structure like PCs. After con-
nection to the network and compatible Bender products, all system devices can be accessed from any PC using
a web browser. In this way, all important system information is directly available. Verified web browsers: Mi-
crosoft IE, Mozilla Firefox, Google Chrome
The possibility of integrating all technical equipment into a single panel creates a "technical monitoring centre"
in the respective room. Each panel is individually manufactured and tailored to the requirements of the user.
3.6 Software products used
CP9… devices are equipped with the COMTRAXX® user interface. It is described separately in the "COMTRAXX"
manual.
3.7 Functional description
3.7.1
Interfaces
CP9… devices communicate with the devices and systems assigned via various interfaces:
Internal BMS bus (RS-485) for Bender systems such as EDS46x/49x, RCMS46x/49x and MEDICS. CP9…
devices can be operated as a master or as a slave. When operated as a master, requests are answered
more quickly. The devices can only be operated on the internal BMS bus.
BCOM for new and future Bender systems, such as ISOMETER® iso685-D.
Modbus RTU (RS-485) when operated as a master for Bender universal measuring devices PEM..3 and
also PEM..5 with restricted functionality (full functionality of PEM..5 only via Modbus TCP).
Modbus TCP (Ethernet) for Bender universal measuring devices PEM..5
3.7.2
Process image
The CP9… alarm indicator and operator panel combines the information from the different interfaces and ma-
kes it available for operation and visualisation via the web user interface of a PC.
It acts as a central user interface. In order to be able to identify them, one individual address is available for each
device on this interface. BMS, BCOM and Modbus RTU devices receive the appropriate address for their inter-
face. A virtual address is assigned to Modbus TCP devices.
3.7.3
List of compatible BMS devices
3.8 BMS side of the CP9… display an operator panel
The majority of Bender devices communicate via the internal BMS bus.
The CP9… alarm indicator and operator panel can be operated as a master or as a slave.
The CP9… alarm indicator and operator panel is to be operated as a master if
- Parameters are queried or changed
- Certain control commands are issued
Please note that not all BMS masters can surrender their master function!
Содержание B95061094
Страница 8: ...Important information 8 CP9xx Kit_D00399_00_M_XXEN 07 2019...
Страница 10: ...Safety instructions 10 CP9xx Kit_D00399_00_M_XXEN 07 2019...
Страница 14: ...Product description 14 CP9xx Kit_D00399_00_M_XXEN 07 2019...
Страница 22: ...Installation connection and commissioning CP9xx Kit 22 CP9xx Kit_D00399_00_M_XXEN 07 2019...