23
B. Setting the Count mode
The QuadCount™ provides three counting modes (Quick/Normal/Precise mode) according to the
counting area. The QuadCount™ is designed to capture multiple frames per channel using an XYZ
stage. Each single image frame covers a volume of 0.15 µL. The more pictures taken, the higher the
accuracy of results.
Select the count mode depending on requirements, refer to the following table.
Count mode
the number of
frames captured
per a channel
Analyzed
volume
Counting
time
per
chamber
Application requirement
Quick mode
1
0.15µL
≤
20s
When you want to get a result quickly and
make a rough estimate of cell numbers.
Normal mode
(default)
6
0.9µL
≤
30s
When you want to get results with
reasonable accuracy and speed (such as
general subculture procedure)
Precise mode
24
3.6µL
≤
100s
When you require precise results or count
cells from a low concentration sample.
NOTE: if the cell concentration is less than 5X10
4
cells/ml, Precise mode is recommended.
STEP 1.
Press the Count Mode button.
2