[ 2 ] PRIORITY OF SELECTIONS
Factory default setting :
Cone
:
Short
cone
Film Speed
:
off
Digital Imaging
:
"d.10"
kV selection
: 60 kV
mA selection
: 6 mA
Patient Type
: Adult
If necessary, these settings can be changed.
For example, if
"D-speed" film
and long cone with
rectangular collimator is used for pedodontistry, default settings for
"Film"
, "Cone"
and "
P
atient
"
1. Keep kV selection switch and mA selection switch depressed simultaneously for more than
3 seconds. Release the switches if the ready light starts to flash.
2. Press
F
switch momentarily.
(Light
for
"Film a"
illuminates and speed setting for
"Film a" is
displayed on exposure time display window.)
3. Select the patient type "child" by depressing P switch momentarily.
5
. Press
T1 switch
until buzzer beeps twice
to store these settings, then turn the main power switch
off.
6
. Cone type, kV and mA selection can be changed by same procedures.
[ 3 ] ELECTRONIC CHIME ON/OFF
An electronic chime sounds when switches are depressed. If preferred, this sound can be
deactivated as follows :
1. Keep T1 and T2 switches depressed together for more than 3 seconds.
Release the switches if the ready light starts to flash.
2. "bu. 2" will be displayed in exposure time display window.
3. By depressing either or switch, display changes to "bu.0".
4. Press
P switch
(Patient type Switch)
until the buzzer beeps twice
to store this setting and turn
off the main power switch.
NOTE
: Exposure Warning Buzzer and alarm sound of error code can not be eliminated.
[ 4 ] ESTIMATED AIR KERMA DISPLAY SETTING
With factory setting the estimated air kerma can be displayed only when the patient type
selection switch is depressed more than 1 second. If automatic display after each exposure
is prefered, change the display setting as follows.
1. Keep T2 and T5 switches depressed together f
o
r more than 3 seconds.
Release the s
w
itches if the ready light starts to flash.
2. "rd.1" will be displayed in exposure time display window.
3. By depressing either or switch, change display to "rd.2".
4. Press P switch (Patient type Switch) until the buzzer beeps twice to store this setting and turn
off the main power switch.
NOTE
: If "rd.0" is stored, estimated air kerma can
not
be displayed automatically
n
or manualy.
-
27
-
should be changed as follows.
4
.
Press C switch several times until the light for long cone flashing. (Flashing means rectangular
collimator is attached to the cone.)