[ 4 ] CONFIRMATION OF

 KV and MA

    

1. Turn the main switch on and set the exposure time at 1 sec. and 60 kV, 3 mA.  

   

 2. Make an exposure and keep the exposure switch depressed continuosly after the exposure is over.

    3. Keeping the exposure switch depressed, press kV selection switch twice.  KV measured value

        

will be displayed on the LED window.  This value should be 60

±

5 kV.

    4. Keeping the exposure switch depressed, press mA selection switch twice.  MA measured value 

        will be displayed on the LED window.  This value should be 3

±

1 mA.

    5. Release the exposure switch and change the setting to 70 kV, 6 mA.
    6. Make an exposure and keep the exposure switch depressed continuosly after the exposure is over.
    7. Keeping the exposure switch depressed, press kV selection switch twice.  KV measured value
        will be displayed on the LED window.  This value should be 70

±

5 kV.

    8. Keeping the exposure switch depressed, press mA selection switch twice.  MA measured value
        will be displayed on the LED window.  This value should be 6

±

1 mA.

 

  

[ 5 ] CONFIRMATION OF EXPOSURE WARNING LIGHT & BUZZER

   

A. EXPOSURE WARNING BUZZER

1. Make an exposure and confirm that the exposure warning buzzer located within the sub controller

        is activated during the entire exposure.

   B. EXPOSURE WARNING LIGHT

       Exposure warning light is located on the front panel of the sub controller,

1. Make an exposure and confirm that the warning light illuminates during the exposure.

 [ 

6

 ] CONFIRMATION OF LINE VOLTAGE REGULATION

1. Make sure that main power switch is "OFF".
2. Set the range of digital multimeter at 

3

00 VAC, connect probes of multimeter to L and N of the

3P terminal block in the main controller.

Do not touch the restriction plate (refer to Fig.3-3) with the probes of multimeter during

       measurement, or a short circuit occurs.

3. Turn the main power switch on, and set the exposure time at 2.00 sec. with manual switch at

70kV, 

6

mA.

4. Record the no-load line voltage (VN) indicated by the multimeter before exposure.
5. Make an exposure and record the load voltage (VL) indicated by the multimeter during exposure.

X-Radiation is generated for 2 seconds.

NOTE :

Read the multimeter when the value is stabilized (about one second after exposure started).

6. Calculate line voltage regulation R(%) in the formula below  :

                                           

NOTE  :

Line voltage regulation must not exceed the range of 0 ~ 5 % for 100, 110, 120V and 0 ~ 3

                 % for 220, 230, 240Vac.  If it is greater than this range, the size of the power supply wires 

                   must be increased.  Refer to the power supply  requirements outlined on page 5 to determine
                   the correct wire size necessary.   If line voltage regulation is within the range, apparent
                   resistance of supply line can be  considered to be in the range of value specified on page 2.

 

 

WARNING

WARNING

P;v/..

TL+TJ
TJ

25

Содержание PHOT-X IIs 505

Страница 1: ...nstallation assembly calibration and certification procedures for BELMONT PHOT X IIs 505 dental x ray The instructions contained in this book should be thoroughly read and understood by dealer service...

Страница 2: ......

Страница 3: ...IDENTIFICATION AND OPERATION 1 MAJOR COMPONENTS AND CONTROL DENTIFICATION 17 2 FUNCTION OF CONTROLS 18 3 OPERATING PROCEDURES 21 4 ESTIMATED AIR KERMA 21 5 OPTIONAL HAND EXPOSURE SWITCH 21 6 ERROR CO...

Страница 4: ...utput 60 kV 70 kV 3 mA 6 mA 3 mA 6 mA a Distal end of regular cone 4 6 9 1 5 9 11 8 mGy sec 40 b Distal end of long cone 2 0 4 1 2 6 5 2 mGy sec 40 Data obtained by direct measurement in the useful be...

Страница 5: ...Maximum Reach 185 1 15 PHOT X IIs Sub Controller 34 mm Front to Back Box Thickness P 150 178 211 7 7 23 100 6 1 130 1 16 0 1 29 4 85 2 31 50 973 1023 1243 Short Cone Long Cone Cabinet Frame for X ray...

Страница 6: ...both B Duty cycle A cool down interval of 30 seconds or more must be allowed between each 1 second exposure a 15 second cool down must be allowed between each 0 5 second exposure This will avoid the a...

Страница 7: ...n device Use a flexible cable approved by CEE 13 52 or 53 consists of 0 75 mm2 or 1 mm2 conductors Diameter of the sheath of cable should be 6 7 5 mm diameter Line voltage regulation should be within...

Страница 8: ...wing radiation requirements concerning the operator s positioning must be considered The operater must 1 have full view of the patient 2 have means for audio and visual communications with the patient...

Страница 9: ...ENTS 2 UNPACKING Unpack the entire contents of the shipping carton Included within the shipping carton are Hardware Quantity Head with Regular Cone 1 Main Controller 1 Screw for chassis M4 x 10 mm 4 S...

Страница 10: ...ring matches to the access hole of wall plate template 2 Mark the hole locations for the arm mounting bracket lag screws 3 Use a 3 16 5mm dia drill to make a pilot hole approximately 50mm deep for eac...

Страница 11: ...nting bracket Fig 3 5 Insert the cable of horizontal arm into the hole and mount to the arm mounting bracket as shown in Fig 3 5 2 Insert two retaining bolts into the upper threaded holes of the arm m...

Страница 12: ...de toward the wall to close Fig 3 9A 2 Install two M3 x 8mm screws into the top of the cover and confirm that the cover is securely attached Fig 3 9B M3x8 Screw Front Cover Front Cover Front Cover Cha...

Страница 13: ...3A Note Two ring terminals should not protrude from a chassis plate Fig 3 13A 5 Secure the wires from the balance arm to the bottom cover with the nylon cable clamp to prevent damage from twisting Fig...

Страница 14: ...n the band 1 Remove the arm collar screw M4 x 8mm from the arm collar Slide the arm collar upward and temporarily hold it in position with adhesive tape Fig 3 15 2 Open the yoke inside cover of x ray...

Страница 15: ...ith the nylon cable clamp Fig 3 18 7 Reattach the yoke inside cover with the screw M4 x 15mm Before closing the cover note the CP values on the CP values label inside of the yoke Fig 3 16 8 Remove arm...

Страница 16: ...ler Fig 3 21 Miswiring causes permanent damage to both timer PC board and power PC board 5 If wire length is too long push it back into the access hole of the wall This will prevent mechanical damage...

Страница 17: ...ith the tapping screw 3 x 8mm Fig 3 25 Fig 3 23 Hand Exposure Switch Kit Hand Exposure Switch Hook Tapping Screw 3 x 8 Fig 3 24 Connecting Hand Exposure Switch Timer PC Board Front Panel CN3 CN2 CN3 H...

Страница 18: ...light tendencies to drift can be corrected by tightening brake screw in horizontal arm Do not tighten beyond what is required to prevent drift 2 BALANCE ARM TENSION ADJUSTMENT 1 Place the balance arm...

Страница 19: ...tion Switch 2 Ready Light 11 Film Speed Selection Switch 3 ExposureTimeAdjusting Switch Down 12 Digital Imaging Switch 4 ExposureTimeAdjusting Switch Up 13 kV Selection Switch 5 Tooth Selection Switch...

Страница 20: ...GHSUHVVHG PRUH WKDQ VHF XQLW JRHV LQWR RFN 0RGH Q ORFN PRGH WKH RQO IXQFWLRQDO VZLWFK LV WKH SRZHU VZLWFK 7R H LW IURP WKH ORFN PRGH GHSUHVV WKH 7 VZLWFh PRUH WKDQ VHF DJDLQ RQH 7 SH 6HOHFWLRQ 6ZLWFK...

Страница 21: ...Q WKH SRVXUH 7LPH LVSOD LQGRZ LWK WKH IDFWRU VSHHG VHWWLQJ G 0 WKH H SRVXUH WLPH EHFRPHV KDOI RI 10 VHWWLQJ As the sensitivity is different according to each manufacturer of digital imaging sensors th...

Страница 22: ...13 0 16 0 18 0 22 0 32 0 22 0 25 0 32 0 36 0 50 0 28 0 32 0 40 0 45 0 63 3 0 18 0 22 0 28 0 32 0 45 0 32 0 36 0 45 0 50 0 71 0 40 0 45 0 56 0 63 0 90 6 0 09 0 11 0 13 0 16 0 22 0 16 0 18 0 22 0 25 0...

Страница 23: ...ures when the unit is not in XVH 127 I WKH XQLW OHIW RYHU PLQ ZLWKRXW EHLQJ RSHUDWHG DQG WKH 0DLQ 3RZHU 6ZLWFK LV NHSW RQ ILJXUH UXQV WKURXJK WKH SRVXUH 7LPH LVSOD LQGRZ 7KLV GRHV QRW PHDQ WKDW PDOIXQ...

Страница 24: ...re than 4 5 mA at 3mA setting or more than 9 mA at 6 mA setting Setting for pre heating time is out of range Tube current is detected during pre heating period Tube current is detected when main power...

Страница 25: ...QLQJ ZLWKLQ WKH PDQXIDFWXUH V VSHFLILFDWLRQV DQG UHPDLQV LQ FRPSOLDQFH ZLWK WKH 6WDQGDUG W LV UHVSRQVLELOLW RI WKH RZQHU RI WKH XQLW WR VHH WKDW WKHVH PDLQWHQDQFH FKHFNV DUH correctly performed Any pa...

Страница 26: ...re making an exposure check this value to be same as the value stored in the subcontroller 1 Confirm the tube potential compensation CP values for 60kV and for 70kV written on the label attached insid...

Страница 27: ...g light is located on the front panel of the sub controller 1 Make an exposure and confirm that the warning light illuminates during the exposure 6 CONFIRMATION OF LINE VOLTAGE REGULATION 1 Make sure...

Страница 28: ...ase film speed number until desired number for a setting is displayed 3 To change the b setting from the factory default F 05 push F switch momentarily until the b light illuminates By depressing or s...

Страница 29: ...conds Release the switches if the ready light starts to flash 2 bu 2 will be displayed in exposure time display window 3 By depressing either or switch display changes to bu 0 4 Press P switch Patient...

Страница 30: ...GE BLUE BROWN BLACK GRAY WHITE ORANGE PURPLE YELLOW BLUE BROWN GRAY RED WHITE BLUE GRAY RED WHITE YELLOW BROWN BLACK BROWN GREEN YELLOW RED BLUE YELLOW BLACK 3 1 2 SWITCH EXP 1 2 3 4 TIMER PC BOARD RE...

Страница 31: ...racket to the wall plate are loose find the reason why bolts became loose take counter measure that prevent bolts become loose 6 Dosimetry Save the image that was taken under appropriate conditions as...

Страница 32: ...ats Mats Mats Mats APPENDIX 3 PARTS IDENTIFICATION 1 CONTROLLER ASSEMBLY C T5 T2 T3 T4 kV 60 mA INC CUS PM CUS PM MLR OCC MLR BW 31 33 34 39 32 44 27 35 29 30 28 43 Mats 36 11 11 40 41 45 46 1 2 38 3...

Страница 33: ...r PC Board 240V Timer PC Board 110V Timer PC Board 100V Timer PC Board 120V Timer PC Board 220V Timer PC Board 230V 31 ECQR68B0 Hook for Hand Exposure Switch Option 32 Hand Exposure Switch Hand Exposu...

Страница 34: ...e Inside Cover Screw M4 x 15 1 5 Head Housing Cover Set 1 set 6 Yoke Side Cap 1 7 Yoke Side Cap Screw 3 x 8 1 8 Lock Ring 1 9 Regular Cone 1 10 Long Cone Option 1 11 Rectangular Collimator Option 1 12...

Страница 35: ...NOTE...

Страница 36: ...Book No 1A0GDSG0 Printed in Japan 2017 04 2 1 1 Higashishinsaibashi Chuo ku Osaka 542 0083 Japan TEL 81 6 6213 5945 TELEFAX 81 6 6212 3680 TAKARA BELMONT CORPORATION NOTE R...

Отзывы: