-12-
7. Insert the brake plug and brake screw (M6 x 6mm)
into the horizontal arm collar. (
Fig.3-14
)
Do not fully tighten.
8. Remove the end cap from horizontal arm.
Insert the stopper screw into upper threaded hole
inside horizontal arm and tighten securely.
Replace the end cap. (
Fig.3-14
)
If stopper screw is not tightened securely, the
Balance Arm can lift out of the horizontal arm.
[ 4 ] HEAD ASSEMBLY INSTALLATION
Do not release Arm holding band until the
X-ray head has been installed.
Balance arm assembly is spring loaded and
can cause equipment damage and injury
if not handled in the proper manner.
Refer to the Caution Tag on the band.
1. Remove the arm collar screw (M4 x 8mm) from
the arm collar. Slide the arm collar upward and
temporarily hold it in position with adhesive tape.
(
Fig.3-15
)
2. Open the yoke inside cover of x-ray head by
removing (M4 x15mm)
cover
screw.
(
Fig.3-16
)
CAUTION
WARNING
Fig.3-14
Attaching Balance Arm
to Horizontal Arm
Fig.3-15
Setting Arm Collar on
Balance Arm
CAUTION !
DO NOT RELEASE
THIS BAND UNTIL
X-RAY HEAD IS
INSTALLED
CAUTION !
DO NOT RELEASE
THIS BAND UNTIL
X-RAY HEAD IS
INSTALLED
Arm Holding
Band
Caution Tag
Adhesive
Tape
Arm Collar
Screw
(M4 x 8)
UP-Mark
Arm
Collar
UP
M
ats
M
ats
Fig.3-16
Removing Yoke Inside Cover
M
ats
Yoke
X-ray Head
CP Values Label
Cover
Yoke inslde
Cover Screw
(M4 x 15)
Screw Driver
Mats
M
a
ts
Brake Screw
Horizontal
Arm
Horizontal
Arm Collar
Stopper
Screw
Arm End Cap
Brake Plug
Balance Arm