Hyperthermia Pump: Parameters Setting and Preventive Maintenance
Page 6 of 20
B.
SERVICE AND PREVENTIVE MAINTENANCE SCHEDULE
Schedule 1
To be performed by either the Clinical User or a Biomedical Technician (BMET).
Routine Maintenance
Interval
Before or After
Each Use
Every
Month
Every 6
Months
1.
Clean and/or Disinfect Exterior, if necessary.
•
2.
Clean Fluid Out and In-Line Air Detector.
•
3.
Check the Power Cord.
•
4.
Clean Temperature Probes
•
5.
Check/Clean the Fan Guard.
•
6.
Check/Clean the Vacuum Trap
•
7.
Check the System Seal.
•
8.
Check Instrument Door and Ceramic Disk.
•
9.
Check Rubber Feet.
•
Schedule 2
To be performed by either a BMET or other qualified service personnel.
Required Test/Verification
Interval
Every 6
Months
Every
Year
1.
Perform Visual Inspection.
•
2.
Perform System Operational Check-Out, including the
Audible Alarm Test.
•
3.
Perform Electrical Safety Test.
•
4.
Hardware Verification.
•
5.
Clean Pump Head
•