17
External choice
By selecting the "external choice" calibration
mode, you can calibrate the scale with a different
weight than the default one.
After confirming, the calibration procedure will be
activated.
Load on the plate a weight equal to or greater
than the default calibration weight, so the scale
will recognize as valid a weight equal to or
greater than the calibration weight as long as it is
a full value weight with respect to the most
significant figure of the default calibration weight.
Example: if the calibration weight is 20g, you can
calibrate the scale with values ranging from 20g,
40g, 60g up to the upper limit of scale capacity.
Load the weight or press the key
to
cancel the calibration procedure.
Wait for the weight to be acquired.
If the operation is successful, the weighing
screen will be displayed with the value of the
calibrated weight.
Remove the weight from the plate.
5.6 Display and print of the scale calibration data.
This section describes how to view and print the calibration data.
Press the menu key
to access the main menu.
Use the keys to move up and down in the menu
and go to section
3 – Scale Cal. Data.
Then confirm the selection with the key
.
Calibration Data:
in the calibration data screen,
you can check the date on which the last
calibration was performed, the mode in which it
was performed, the value of the weight used and
the correction made with respect to the previous
calibration. Press the key
to print the
displayed data.
Press the key
to exit.