16
NOTE: an error message will be displayed if there is some interference during the calibration process.
To interrupt the calibration process, press the ON/OFF button while the calibration weight indication
flashes.
It is also possible to calibrate the balance with a calibration weight greater than the pre-set
calibration weight:
1. Press and hold the
CAL
button with the plate empty until the beeping stops, and then
release the button. The word
“-CAL-“
will be shown on the display, followed by the word
“LOAD”
, flashing.
2. Load a weight that is equal to or greater than the pre-set calibration weight on the plate;
the balance will recognize a weight that is equal to or greater than the calibration weight
as valid provided that it is a whole weight with respect to the most significant figure of the
calibration weight.
E.G.:
if the calibration weight is 200 g, it will be possible to calibrate the balance with
values that go from 200g, 300g, 400g up to the upper capacity limit of the balance.
The word
“LOAD”
on the display will stop flashing; once the calibration has been carried
out, the value of the weight used will be displayed.
3. Remove the calibration weight; the balance is ready for the weighing operations.
2 Sec
Cal
01-06-10 13:15:24
-0-
Load
01-06-10 13:15:24
-0-