7.  MAINTENANCE

7.1 Fault finding during commissioning

If a BA544D fails to function during
commissioning the following procedure should
be followed:

Symptom

Cause

Check:

No display

Instrument has
not been
switched ON.

Battery is not
conditioned,

Position of slide
switch above
terminals 3 & 4.
See Figs. 3 & 4.

See Appendix 1

Flow indicator

not rotating

No input
pulses,
incorrect input
terminals
used, or
incorrect type
of input
selected

That BA544D has
been calibrated for
the correct input.

That input signal is
connected to
correct input
terminals and
correctly polarised.

Flow indicator

rotating but

incorrect rate

display

Incorrect
calibration

Input below
clip-off
threshold

That rate scale
factor and
timebase are
correctly
programmed.

That clip-off
threshold is set
correctly.

Unstable rate

display

Noisy pulse
input signal

Source of noise
and try to
eliminate, or
increase rate filter.

Correct rate

display but

incorrect total

display

Incorrect
calibration

Remote reset
switch
contacts
closed

That total scale
factor is correctly
programmed.

That ‘RESET’
annunciator is not
activated.  If it is,
check reset wiring
and switch.

Unable to

enter

programme

mode

Incorrect
security code

That the correct
security code is
being used, or fit
security link in
override position.
See Fig 8.

7.2  Fault finding after commissioning

ENSURE PLANT SAFETY BEFORE

STARTING MAINTENANCE

If a BA544D fails after it has been functioning
correctly, the following table may help.

Symptom

Cause

Check:

No display

Instrument has
been switched
OFF

If BA544D has
been switched
OFF for
extended
period battery
may need
conditioning.

Discharged
battery

Position of slide
switch above
terminals 3 & 4.
See Figs. 3 & 4.

See Appendix 1

f replacement
battery restores
operation

Low battery
annunciator

activated

Battery almost
exhausted.
Install
replacement
battery as
soon as
possible.

Instrument

displaying

‘bAtt’

Battery
exhausted.
Install
replacement.

Flow indicator

not rotating

No input
pulses

Output from
flowmeter and
wiring between
flowmeter and
BA544D.

Flow indicator

rotating,

‘HOLD’

annunciator

activated

Input below
clip-off
threshold

Clip-off threshold
and if necessary
adjust.

Unstable rate

display

Noisy pulse
input signal

Source of noise
and try to
eliminate, or
increase rate filter.

If this procedure does not reveal the cause of
the fault, it is recommended that the instrument
is replaced.

7.3  Servicing

We recommend that faulty BA544D rate
totalisers are returned to BEKA associates or
to our local agent in your area for repair.

7.4 Routine maintenance

The mechanical condition of the instrument
and electrical calibration should be regularly
checked. The interval between inspections
depends upon environmental conditions.  We
recommend that initially instrument calibration
should be checked annually.

7.5  Guarantee

Instruments which fail within the guarantee
period should be returned to BEKA associates
or our local agent.  It is helpful if a brief
description of the fault symptoms is provided.

Содержание BA544D

Страница 1: ...Issue 3 7th April 2009 BA544D Battery powered pulse input field mounting rate totaliser Issue 3...

Страница 2: ...total 5 15 Security code 6 Calibration Example 6 1 Calibration procedure 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routine maintenanc...

Страница 3: ...further extend the applications of this instrument The BA544D contains an internal replaceable battery that will power the instrument for up to five years Battery life depends upon how the BA544D is p...

Страница 4: ...ue to operate for about five hundred hours after the annunciator is activated but the exact time will depend upon how the instrument is programmed whether alarms are fitted and the input frequency Dur...

Страница 5: ...ater treatment installations Please consult BEKA associates if high vibration is anticipated The BA544D enclosure is surface mounting Accessory kits described in sections 8 4 8 5 of this manual enable...

Страница 6: ...one point Fig 4 Dimensions and terminal connections 4 4 Installing a replacement battery The internal battery will power the BA544D rate totaliser for about five years the exact time depends upon how...

Страница 7: ...by moving the slide switch to the left see Figs 3 4 c Remove the four D screws and lift off the front of the enclosure d Remove the two E screws securing the BA491 battery and gently pull the battery...

Страница 8: ...ond minute or hour Similarly to calculate the total flow in the required units of measurement the number of input pulses is divided by the total scale factor SCALE t Total flow is continuously added t...

Страница 9: ...9...

Страница 10: ...section 5 6 dISPLAY 2 Lower display Turns the lower display which normally shows rate on or off See section 5 7 Display Description of function SCALE r Rate Scale Factor Defines the arithmetic relatio...

Страница 11: ...ype select InPut from the menu and press P which will reveal the current input type Pressing Up or Down will scroll through the four options UOLts Voltage pulse input Threshold 1V and 3V 10kHz max Inp...

Страница 12: ...int position and simultaneously press the Up and Down buttons When the required rate scale factor has been entered press E to store the number and return to the main menu 5 9 Timebase t bASE The timeb...

Страница 13: ...he total scale factor the total and grand total displays will be reset to zero To position the decimal point in the scale factor move the flashing digit to the left hand side of the required decimal p...

Страница 14: ...ed before access to the programme menu is granted New instruments are programmed with the default security code 0000 which allows unrestricted access to all programming functions without entering a se...

Страница 15: ...Select the type of input Using the Up or Down button scroll through the main menu until InPut is displayed then press P to reveal the existing setting Select COIL the input for a magnetic pick off usi...

Страница 16: ...A544D is required to display 1 cubic metre for every 105 x 1 000 pulses received SCALE t should therefore be set to 105 000 Select SCALE t from the main menu and press P which will reveal the current...

Страница 17: ...rument to the CLr Gtot prompt in the main menu Step 12 Define the security code Defining a security code prevents unauthorised access to the programme functions Select COdE from the main menu and pres...

Страница 18: ...fter it has been functioning correctly the following table may help Symptom Cause Check No display Instrument has been switched OFF If BA544D has been switched OFF for extended period battery may need...

Страница 19: ...d not be used for critical safety applications such as a shut down system When the battery is exhausted both BA544D alarm outputs will open irrespective of whether normally open or normally closed out...

Страница 20: ...of alarm Defines whether the alarm operates on the rate or total display See section 8 2 4 SPr1 Alarm setpoint 1 or Adjusts the alarm setpoint The SPt1 alarm is activated when the rate or total displa...

Страница 21: ...rn to the alarm menu 8 2 4 Type of alarm tYPE Alarm 1 and Alarm 2 may both be rate or total alarms or one may be conditioned for rate and the other for total To check or change the type of alarm selec...

Страница 22: ...s select HStr from the alarm menu and press P which will reveal the existing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next digit When the re...

Страница 23: ...o the setpoints in the display mode by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with the security code set to 0000 8 2 12 Adjusting alarm setpoints from th...

Страница 24: ...mounting kits are available for securing the BA544D to a horizontal or vertical pipe BA392D Stainless steel bracket secured by two worm drive hose clips for 60 to 80mm outside diameter pipes BA393 He...

Страница 25: ...Y 2 5 7 d P 5 6 EnbL 8 2 2 8 2 3 Fault finding during commissioning 7 1 after commissioning 7 2 FILtEr 5 10 Flow indicator 2 2 Grand total 2 2 clear 5 14 Guarantee 7 5 HI LO 8 2 2 8 2 6 HStr 8 2 2 8 2...

Страница 26: ...t is switched on When a replacement battery is fitted or when the rate totaliser is turned on after not being used for an extended period the battery should be conditioned by drawing a relatively larg...

Отзывы: