background image

8.  ACCESSORIES

8.1  Scale card

Both models have a window on the right hand side of
the  display  through  which  to  view  a  scale  card
showing the units of measurement such as 

o

C, mBar,

RPM.    New  indicators are  fitted  with  a scale  card
showing the units of measurement specified when the
indicator was ordered, if the units are not specified a
blank scale card will be fitted.   A pack of scale cards
pre-printed  with  common  units  of  measurement  is
available as an accessory.  These can easily be fitted
on-site to the indicator without opening the indicator
enclosure or removing it from the panel,  See section
4.5 of this instruction manual.

Custom  scale  cards  for  applications  requiring  less
common units of measurement are also available.

8.2  Tag information

New  indicators  are  supplied  with  tag  or  application
information laser etched onto the rear panel adjacent
to  the  terminals,  legend  as  specified  when  the
indicator  was  ordered.    This  tag  information  is  not
visible  from  the  front  of  the  instrument  after
installation.

8.3   Alarms

CAUTION

These  alarms  outputs  should  not  be  used  for
critical safety applications such as an emergency
shut down system.

Both models can be supplied with factory fitted dual
solid state, single pole alarm outputs.   Each alarm
output may be independently conditioned as a high or
low  alarm  with  a  normally  open  or  normally  closed
output in the non-alarm condition.

When  the  4/20mA current  powering  the  indicator  is
removed both alarm outputs will open irrespective of
configuration.    The  open  circuit  condition  should
therefore  be  chosen  as  the  alarm  condition  when
designing an alarm system.    Fig 11 illustrates the
conditions available and shows which are fail safe.

When an alarm occurs an alarm annunciator on the
indicator front panel is activated and if required the
numerical  display  can  alternate  between  the
measured value and the alarm channel identification
‘ALr1’ or ‘ALr2’.

CAUTION

The   alarms   are   activated   by   the   indicator’s
numerical  display.    Use  of  the  Tare  Function
‘tArE’  will  change  the  numerical  display,  the
alarms  will  continue  to  function  at  the  original
displayed  value,  but  this  will  correspond  to  a
different input current.

Fig 11  Alarm outputs

Configurable  functions  for  each  alarm  include
adjustable  setpoint,  hysteresis,  alarm  delay  and
alarm accept.

8.3.1  Solid state output 

Each alarm has a galvanically isolated single pole
solid state switch output which as shown in Fig 12.
The output is polarised and current will only flow in
one direction.

Ron

=

less than 5

 + 0.7V

Roff

=

greater than 1M

Fig 12 Equivalent circuit of each alarm output

The solid state output of each alarm may be used
to switch any dc circuit with parameters equal to or
less than:  

V

=

30V

I

=

200mA

18

Содержание BA507E-SS

Страница 1: ...Issue 2 4th May 2017 BA507E SS BA527E SS Rugged general purpose loop powered panel mounting indicators Issue 2...

Страница 2: ...ints to a new indicator 6 3 Under and over range 6 4 Lineariser default configuration 7 Maintenance 7 1 Fault finding during commissioning 7 2 Fault finding after commissioning 7 3 Servicing 7 4 Routi...

Страница 3: ...be impacted The zero and span of the display are independently adjustable so that the indicator can be calibrated to display any variable represented by the 4 20mA input current e g temperature flow...

Страница 4: ...s in display mode function has been enabled P E Access to configuration menu via optional security code Note BA527E SS only If the indicator has been calibrated using the CAL function calibration poin...

Страница 5: ...b Inspect the indicator s captive gasket and ensure that it is not damaged before inserting the indicator into the panel enclosure aperture c If the enclosure panel is less than 1 0mm thick or is non...

Страница 6: ...he panel enclosure in which it is mounted New indicators are supplied with a printed scale card showing the requested units of measurement if this information is not supplied when the indicator is ord...

Страница 7: ...the self adhesive printed scale card onto the flexible strip and insert the strip into the indicator as shown below Fig 7 Inserting flexible strip carrying scale card into slot at the rear of indicat...

Страница 8: ...Function FunC Std Std Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00 Bargraph finish BarHi 100 00 P button in display mode C...

Страница 9: ...9...

Страница 10: ...lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or press the or button to cha...

Страница 11: ...A If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the instrument will be reverse acting i e an i...

Страница 12: ...between on and oFF When set as required press E to return to the tARE prompt in the configuration menu 5 10 Security code CodE Access to the instrument configuration menu may be protected by a four di...

Страница 13: ...ct calibration of the lineariser with an external current source and is the preferred method when traceability is required If the exact system non linearity is unknown this method also allows direct c...

Страница 14: ...14...

Страница 15: ...wing break point 6 1 1 Example Adding break points to a new indicator When adding break points to a new indicator or to a lineariser following resetting to the factory defaults using the LtAb function...

Страница 16: ...e corresponding indicator display at each of the break points can now be defined using the diSP sub function Using the and buttons select the diSP sub function and press P which will reveal the starti...

Страница 17: ...tion 7 1 If this does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20m...

Страница 18: ...ut may be independently conditioned as a high or low alarm with a normally open or normally closed output in the non alarm condition When the 4 20mA current powering the indicator is removed both alar...

Страница 19: ...19...

Страница 20: ...SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator display equals the setpoint See section 8 3 4 Hi Lo Alarm function Defines the alarm function as High or Low...

Страница 21: ...fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press P...

Страница 22: ...g access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct set...

Страница 23: ...ayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 16 Fig 16 Displayed value and setpoints on bargraph The bargraph ar...

Страница 24: ...17 Terminals for optional backlight 8 4 1 Loop powering the backlight The backlight is loop powered by connecting it in series with the indicator 4 20mA input as shown in Fig 18 which increases the ma...

Отзывы: