Safety
Operating manual
KINET
V1.4 E
Page 9 of 40
4.5 Warnings
Note
▪
The product may only be operated by trained staff.
▪
Protect the product against direct sunlight and heat.
▪
The equipment is not authorised for use in potentially explosive atmospheres.
▪
Never exceed the duty cycle or the maximum load.
▪
Never cover up, oversticker or change the slots and holes of the device.
▪
Do not operate or store moisture-sensitive device in the vicinity of the bath tub.
▪
Please ensure that the power supply is always switched on throughout the
treatment.
▪
The operator must ensure that the area above and below the bath tub is free of
obstacles that could obstruct the free tilting movement of the bath tub.
▪
In order to prevent any electromagnetic interference (EMC), the simultaneous
use of short-wave or microwave therapy device in the immediate vicinity of the
device is strictly forbidden. Mobile telephony could also cause interference.
▪
Route the mains and the connecting cable so that they cannot be damaged.
Damaged mains cables could cause fire or lead to electrocution and must not
be used.
▪
The caregivers must protect their skin and eyes against concentrated
disinfecting and cleaning products. Use a face mask to protect yourself against
aerosols.
▪
Check prior to each use that all visible parts are intact. The product must not be
used if any parts are damaged. Prior to each use of the device and its
accessories, the user must check the functional safety and the good condition
of
the device and its accessories (visual check, functioning).
▪
The product must be disinfected after each treatment.
▪
Supervision of the caregiver is required throughout the treatment.
▪
Please make sure that your feet are not located beside the bath tub before and
during the tilting movement.
▪
Do not stand between the bath tub and an obstacle during the transport
procedure.
▪
Use only original accessories for the
KINET
tilting bath tub.
▪
Check every month the pressure hoses for leakage, the connections for
tightness and the electrical connections (current and equipotential bonding)
for cable cracks.
▪
Repairs to components of the
KINET
tilting bath tub are to be carried out
only by trained expert personnel. Please contact the after-sales service. The
opening the device or other accessories will lead to the expiration of all
guarantee, warranty and liability claims.
Содержание 910150700
Страница 27: ...Appendix Operating manual KINET V1 4 E Page 27 of 40 13 2 Dimensions of the KINET tilting bath tub...
Страница 30: ...Appendix Operating manual KINET V1 4 E Page 30 of 40 13 5 Declaration of Conformity KINET...
Страница 38: ...Appendix Operating manual KINET V1 4 E Page 38 of 40 Type of inspection Date Result Measure Signature...
Страница 39: ...Appendix Operating manual KINET V1 4 E Page 39 of 40 Your notes...
Страница 40: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...