Appendix
Operating manual
KINET
V1.4 E
Page 26 of 40
13 Appendix
13.1 Technical data
Dimensions and weights
- Length reclined position:
156.5 cm
- Width:
84.5 cm
- Height reclined position:
92.0 cm
- Max. stroke height:
20.0 cm
- Height entry position:
22.0 cm/128.0 cm (front/back)
- Entry height:
22.0 cm
- Angle of inclination:
45°
- Weight without packaging:
approx. 90 kg
- Filling capacity (without patient):
max. 178 l
- Filling capacity (with patient):
approx. 80 l - 120 l
-
Patient’s weight:
max. 210 kg
- SWL (safe workload):
max.
325 kg
(patient and water)
Electrical data
- Supply voltage:
230V, 50 Hz
or
120 V, 60 Hz
- Power:
1440 VA
- Protection class:
II
- Applied part:
Type B
Ambient conditions
Operation
- Temperature range:
10° C to 40° C
- Relative humidity:
30% to 95%, non-condensing
- Atmospheric pressure:
800
– 1060 hPa
Storage and transport
- Temperature range:
-20° C to 70° C
- Relative humidity:
10% to 80%, non-condensing
- Atmospheric pressure:
500
– 1100 hPA
Содержание 910150700
Страница 27: ...Appendix Operating manual KINET V1 4 E Page 27 of 40 13 2 Dimensions of the KINET tilting bath tub...
Страница 30: ...Appendix Operating manual KINET V1 4 E Page 30 of 40 13 5 Declaration of Conformity KINET...
Страница 38: ...Appendix Operating manual KINET V1 4 E Page 38 of 40 Type of inspection Date Result Measure Signature...
Страница 39: ...Appendix Operating manual KINET V1 4 E Page 39 of 40 Your notes...