Appendix
Operating manual
KINET
V1.4 E
Page 36 of 40
Recommended distance between portable and mobile communications equipment
and the
KINET
(Table 206)
The
KINET
is intended for use in an electromagnetic environment with controlled HF
interferences. The customer or the user of the
KINET
can avoid electromagnetic interference
by respecting and observing the minimum distance between portable and mobile HF
telecommunication devices (transmitters) and the
KINET
depending on the rated output of the
communication device as given below.
Rated output
of the transmitter
W
Separation distance depending on the transmitting frequency
in m
150 kHz to 800 MHz
d=0.35√P
80 MHz to 800 MHz
d=1.2√P
800 MHz to
2.5 GHz
d=2.3√P
0.01
0.1
1
10
100
0.04
0.11
0.35
1.1
3.5
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in
watts (W) according to the specifications given by the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2:
These guidelines could not apply to all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Содержание 910150700
Страница 27: ...Appendix Operating manual KINET V1 4 E Page 27 of 40 13 2 Dimensions of the KINET tilting bath tub...
Страница 30: ...Appendix Operating manual KINET V1 4 E Page 30 of 40 13 5 Declaration of Conformity KINET...
Страница 38: ...Appendix Operating manual KINET V1 4 E Page 38 of 40 Type of inspection Date Result Measure Signature...
Страница 39: ...Appendix Operating manual KINET V1 4 E Page 39 of 40 Your notes...
Страница 40: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...