Prism Medical Ltd. V1-2014
5.2 Main board control
14
Picture 1:
Required tools:
1 x battery-operated
screwdriver
1 x bit AW 25
Picture 2:
Additional parts:
1 x control
Main board
incl. screws
Picture 5:
Loosen the four
screws from the
distributor unit
(Aerospa).
Picture 4:
Remove the
cover of the
disinfectant
container
and take the
container out
from the open
back wall.
Picture 3:
Remove all cables
from the control.
Picture 6:
Remove the two
screws that hold the
distributor unit. Then
fold the distributor
unit away toward the
bottom.
Picture 7:
Lift the 2 black lines.
Picture 8:
Remove the left and
right screw from the
control housing.