C-4
On dual range Power Supplies, this command selects the LO or HI voltage range. Dual Ranges are 0-135
volts and 0-135/0-270 volt. If this command is omitted from the setup string the unit will default to the
lowest available range. On single range units this command will be ignored.
NOTES:
1. The setup command specifies the voltage and frequency of the AC power Supply output. The AC Power
Supply responds only to the last setup command entry. Its memory does not retain previously entered
setup commands.
2. During EIA-232 remote operation, the unit's front panel VOLTS and FREQ must be fully CCW. If the
controls are not set fully CCW, then the output voltage and frequency will be a sum of the programmed
value and the value set with the front controls. Front Panel OUTPUT switch must be set to OFF position.
Front Panel RANGE switch must be set to LO.
CILL SOFTWARE COMMANDS
CILL COMMANDS
EXPLANATION
FETCH DATA COMMAND:
FTH VOLT <cr><lf><eos>
The AC supply responds to this command by transmitting the RMS
voltage (in volts) measured at its output, in decimal format as follows:
<sp><digit><digit><digit><dp><digit><digit><cr><lf><eos>
FTH CURR <cr><lf><eos>
The AC Supply responds by transmitting the current (in amps)
measured at its output, in decimal format, as follows:
<sp><digit><digit><dp><digit><cr><lf><eos>
FTH FREQ <cr><lf><eos>
The AC Supply response by transmitting the frequency (in hertz)
measured at its output, in decimal format, as follows:
<sp><digit><digit><digit><cr><lf><eos>
CLS :CH0 <cr><lf><eos>
Closes output relay contacts. Connects AC Power Supply outputs to
output terminal block. The setup command specifies the voltage and
frequency. The setup-command must be entered before the CLS :CH0
command.
OPN :CH0 <cr><lf><eos>
Opens output relay contacts. Disconnects output of AC Power Supply
from output connector. Ac Power Supply retains voltage and frequency
assigned by setup command.
RST ACS:CH0
Resets AC Power Supply to quiescent. The output relay opens and error
messages are erased.
Catastrophic error messages are cleared by reading the Power Supply's
response, by use of the Status command.
CNF <cr><lf><eos>
Instructs AC Power Supply to perform internal confidence test. STA
command transmits the message. The AC Power Supply responds with
one of the following:
Pass
<sp><cr><lf><eos>
CILL SOFTWARE COMMANDS (cont)
CILL COMMANDS
EXPLANATION
Fail
Содержание BL20000 Series
Страница 18: ...13 BL20K SERIES REV E 2013 Figure 3 2 Typical BL 20000 Front and Rear Panel Views ...
Страница 31: ......
Страница 32: ...26 BL20K SERIES REV E 2013 THIS PAGE INTENTIONALLY BLANK ...
Страница 50: ...6 1 SECTION 6 DRAWINGS ...
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......
Страница 61: ......
Страница 62: ......
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......