Voltage measurement
Never apply more than 600V AC/DC to the test leads
(instrument). If these threshold is exceeded the result
could be serious personal injury or damage to the in-
strument.
Only use the instrument in circuits fused with max. 16A.
Before switching to a new function, the probes must
always be removed from the UUT (Unit Under Test).
DC voltage measurement
1) Select measurement range "V
" with Range selec-
tion switch (3)
2) Select range "
" with AC/DC switch key (8)
3) Connect black test lead with COM socket and red test
lead with V
Ω
socket
4) Connect test leads to UUT
5) Read the measured value on the display
AC voltage measurement
1) Select measurement range "V
" with Range selec-
tion switch (3)
2) Select range "~" with AC/DC switch key (8)
3) Connect black test lead with COM socket and red test
lead with "V
Ω
” socket
4) Connect test leads to UUT
5) Read the measured value on the display
Resistance measurement
Ensure that the UUT is switched off and that no volta-
ge is present.
1) Select measurement range "
Ω
/•))" with Range selection
switch (3)
2) Connect black test lead with COM socket and red test
lead with "V
Ω
" socket
3) Connect test leads to UUT
4) Read the measured value on the display
18