6.1 Infrared Interface, send Measurement Data
Additionally, all memorised measurement data may
be transferred to the PC (data archive) by means of
an Infrared Interface. For this purpose, the interfa-
ce adapter and the Software "Report-Studio” and In-
terface, available as an optional feature, are requi-
red.
The data transfer procedure is fairly easy:
Apply the Interface adapter to TELARIS
Connect adapter cable to PC (COM 2) via
9pole D-sub plug (RS232).
Call the "Report-Studio” Software (please
refer to user manual "Report-Studio”).
Select menu item "read data” or select the
button "read data from measurement instru-
ment” by mouse click. The message "send
data” appears.
Press key "Send" (26).
Data transfer is carried out. After a few se-
conds all stored data is available for further
processing in the PC.
For further details, please refer to user manual
"Report-Studio".
7.0 Deleting Stored Measurement Data / Display
of All Memory Entries
It is possible to delete data by using key „Clear" (27).
Furthermore, it is possible to display the total num-
ber of measurement value entries on the display:
Display of Total Number of Measurement Value
Entries:
Set measurement range selection switch
(31) to one of the 5 measurements ranges.
Briefly press „Clear” key (27). A short beep
signal is audible.
The total number of all measurement values
that have been saved is now displayed.
Delete Last Stored Measurement Value:
Set measurement range selection switch
(31) to one of the 5 measurement ranges.
Briefly press „Clear” key (27). A short beep
signal is audible and the number of the mem-
orised measurement data is displayed.The
total number of all stored measurement data
is displayed.The display disappears after ap-
proximately 2 seconds.
Now release the key „Clear” (27) to avoid
deletion of all memory entries.
Deleting the last measurement value can be re-
peated as often as required, as long as data is
available within the memory.
Deleting all Stored Measurement Values:
Set measurement range selection switch
(31) to one of the 5 measurement ranges.
Press „Clear” key (27) for approx. 5 seconds.
A beep signals are audible. The number of all
stored measurement values is displayed.
The last stored value is deleted after 2 sec-
onds. After 5 seconds, all values are deleted.
„0“ appears on the display. The memory is
completely deleted, the „M" symbol in the
display disappears.
46
Infrared Interface, send Measurement Data
PTDB90610009.qxd:PTDB9061-02.qxd 14.08.2007 8:43 Uhr Seite 46
Содержание 9061
Страница 2: ...2 PTDB90610009 qxd PTDB9061 02 qxd 14 08 2007 8 42 Uhr Seite 2 ...
Страница 30: ...30 PTDB90610009 qxd PTDB9061 02 qxd 14 08 2007 8 43 Uhr Seite 30 ...
Страница 56: ...56 PTDB90610009 qxd PTDB9061 02 qxd 14 08 2007 8 44 Uhr Seite 56 ...
Страница 82: ...82 PTDB90610009 qxd PTDB9061 02 qxd 14 08 2007 8 44 Uhr Seite 82 ...