28
8:2010
IEC61000-4-
8:2012
30 A/m 50 Hz and
60Hz
ESD
EN61000-4-
2:2009
IEC61000-4-
2:2012
2, 4, 8kV contact 2,
4, 8, 15kV air
Criterion B
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should at least 30%.
Notes:
*A:
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitter, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the iCOquit® is used
exceeds the applicable RF compliance level above, the iCOquit® should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the iCOquit®.
Emissions test
IEC 60601 Test
level
Compliance
level
Electromagnetic environment – guidance
Radiated Emissions
EN55011:2016+
A1:2017
FCC part 15 Subpart
B
30MHz to 1GHz
30MHz to
25GHz
Class B
The iCOquit® is suitable for use in all establishments,
including domestic establishments, including
domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Transmitter
Radiated Spurious
Emissions
ETSI EN 300 328
V2.2.1:2019
30MHz to
12.75GHz
Class B
Output Power
(EIRP)
ETSI EN 300 328
V2.2.1:2019
-10dBW
(100mW) (<20
dBm)
Class B
Occupied Channel
Bandwidth
(Antenna Port)
ETSI EN 300 328
V2.2.1:2019
2.4 – 2.4835
GHz
For non-
adaptive with
eirp < 10dBm
BW < 5MHz
Class B
Содержание iCOquit Smokerlyzer
Страница 1: ...User Manual iCOquit Smokerlyzer ...
Страница 31: ......