DxC Integrated Workstation In-Lab Training Manual
Page 34 of 101
Rev. A (June 2014)
Beckman Coulter Professional Development
DxI Supplies
DxI
Supply
Capacity
Yellow Caution
Red Warning
Notes
Wash
Buffer
2 x 10 L
1 Container Empty
Both Containers
Empty
Red Warning:
Completes testing in
progress, but no new
tests scheduled
Liquid
Waste
DxI 800:
2 x 9.5 L
DxI 600:
2 x 5.0 L
1 Container Full
Both Containers
Full
Red Warning:
Completes testing in
progress, but no new
tests scheduled
Not used on systems
plumbed directly to a
drain
Reaction
Vessels
1700-1800 <1000
<100
Red Warning:
When the vessel
hopper is empty, it
completes testing in
progress, but no new
tests scheduled
Solid Waste
DxI 800:
20 pounds
(9.1 kg)
DxI 600:
10 pounds
(4.6 kg)
10% or less
volume available in
container for
collecting waste
Container is Full
Red Warning:
When the container is
full, it completes testing
in progress, but no new
tests scheduled
Substrate
2 x 130 mL
(1200 tests)
A lot expires in
3 days
An open bottle
expires in 3
days
One
on-board
bottle is empty
Only one bottle
on-board and it
has <60 tests
left
One bottle is
not present or
not recognized
by the system
All lots are
expired
One or both
open bottles
are expired
All
on-board
bottles are
empty
No bottles
present, or
bottles are not
recognized by
the system
Red Warning:
When both on-board
bottles are empty or
expired, the system
completes any in
progress tests, but no
new tests scheduled
Reagent
Packs
50 positions
<10 tests left
3 days or less
to lot expiration
3 days or less
to calibration
expiration
Expired
pack
lot
Expired
open
stability
Expired or not
calibrated
Supplies Required
button turns yellow to
alert operator when
reagent is empty,
expired or not
calibrated