PN 623112BA
3-3
SPECIFICATIONS
INSTRUMENT SPECIFICATIONS
3
Specimen Type
Whole Blood
Anticoagulant
Use the anticoagulant for patient specimens that is recommended for the flow cytometry
application being tested. Follow the specimen tube manufacturer’s recommended procedure
for the volume of anticoagulant to use.
Specimen Size
5-1,000 µL. (Volumes over 100 µL are delivered in multiple aliquots. Maximum aliquot
volume is 100 µL.)
Specimen Tube Size
See
in the Instructions For Use manual for a detailed list of
acceptable closed-vial specimen tubes.
Specimen Cassette Sizes
Three sizes of specimen cassettes for tubes are available for holding:
r
10-13 mm tube cassette for
t
13 X 75 mm specimen tubes
t
3 mL (10.25 X 64 mm) pediatric tubes in 3 mL tube adaptors
t
2 mL (10.25 X 47 or 50 mm) pediatric tubes in 2 mL tube adaptors
t
Sarstedt tubes
Note
: Sarstedt tubes must be set up by your Beckman Coulter Representative.
Do not mix these tubes in the same cassette. Each tube type will go in its own cassette
based on the cassette bar code.
r
16 X 100 mm tube cassette for
t
16 X 100 mm specimen tubes
t
IMMUNO-TROL control tube in its special cassette adaptor
Be sure to label each cassette with the correct bar-code label.
r
16 X 75 mm tube cassette for
t
16 X 75 mm tubes
Cassette type is differentiated by cassette bar code.
See
in the Instructions For Use manual for more information.
See
in the Instructions For Use manual for a detailed list of which
specimen tubes can be used with each specimen cassette size.
WARNING
Risk of operator injury. Risk of biohazardous condition due to specimen tube breakage. Risk of
damage to sample probe. Failure to use the proper bar-code-labeled cassette based on the type of tube can
result in a probe crash on the bottom of the tube, a bent probe, possible tube breakage, and specimen
spillage. Ensure that you use the proper bar-code-labeled cassette based on the type of tube.
Содержание FP 1000
Страница 4: ...PN 623112BA iv REVISION STATUS ...
Страница 12: ...PN 623112BA xii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Страница 30: ...PN 623112BA 3 6 SPECIFICATIONS PERFORMANCE SPECIFICATIONS ...
Страница 32: ...PN 623112BA A 2 LOG SHEETS ...
Страница 34: ...PN 623112BA A 4 LOG SHEETS ...
Страница 36: ...PN 623112BA A 6 LOG SHEETS ...
Страница 38: ...PN 623112BA A 8 LOG SHEETS ...
Страница 48: ...PN 623112BA INDEX 6 INDEX ...
Страница 50: ......