PN 623112BA
2-7
INSTALLATION
TRAINING CHECKLIST
2
Controls and Calibrators
Reagents
Level sense for reagents is not user-selectable and is always enabled.
Level sense for
specimens is user-selectable for random processing mode only and is not available for
batch processing mode.
Level-sense functionality is optimized for cassette types 0-6. Level-sense functionality for
cassette types 7-9 is based on 13 X 75 mm tubes. For best results, use specimen tube types
specified in the manual.
Specimens
System Operation
B
Review how to reconstitute controls and calibrators on the FP 1000.
B
Review how to load controls and calibrators into the control/cal rack, noting to check for
bubbles/foam.
B
Note that the IMMUNO-TROL control tube must be bar-coded and run in a CONTROL cassette.
B
Review the daily quality control procedure.
B
Review how to reconstitute reagents on the FP 1000.
B
Review how to determine whether or not the racks have sufficient reagents to meet the test
requirements.
B
Review how to load reagents into the different reagent racks (Small, Large/Small, Large, Bulk),
noting to check for bubbles/foam.
B
Review how to properly prepare and load Flow-Count fluorospheres, noting to check for
bubbles/foam.
B
Demonstrate how to replace the IsoFlow container and empty the waste.
B
Demonstrate how to replace IMMUNOPREP reagents, VersaLyse Lysing solution, and/or
IOTest 3 Lysing Solution (ammonium chloride).
B
Describe the reagent level-sensing system, associated errors, and how to respond.
B
Differentiate between the reagents that are monitored by the level-sensing system and those that
are tracked by counters.
B
Demonstrate how to restart the reagent counters.
B
Review proper specimen collection, storage, age, and handling. Note that the system does not
have clot detection. Review the need to vent any specimen tube that was not fully filled (has
vacuum) or has been opened after specimen collection.
B
Review the acceptable specimen tube types.
B
Review which type of cassette to use for each specimen tube type.
B
Demonstrate how to apply bar-code labels to daughter tubes.
B
Demonstrate how to place bar-coded daughter tubes in a carousel.
Содержание FP 1000
Страница 4: ...PN 623112BA iv REVISION STATUS ...
Страница 12: ...PN 623112BA xii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Страница 30: ...PN 623112BA 3 6 SPECIFICATIONS PERFORMANCE SPECIFICATIONS ...
Страница 32: ...PN 623112BA A 2 LOG SHEETS ...
Страница 34: ...PN 623112BA A 4 LOG SHEETS ...
Страница 36: ...PN 623112BA A 6 LOG SHEETS ...
Страница 38: ...PN 623112BA A 8 LOG SHEETS ...
Страница 48: ...PN 623112BA INDEX 6 INDEX ...
Страница 50: ......