BeBionic bebionic3 Скачать руководство пользователя страница 1

technical information

Содержание bebionic3

Страница 1: ...technical information...

Страница 2: ...2 technical information...

Страница 3: ...tructions for bebionic Glove 4 0 Connections Compatibility 27 35 4 1 Wrist Options 4 2 Battery Options 4 3 Power Information 4 4 System Connections 4 5 System Compatibility 5 0 bebalance 37 49 5 1 Int...

Страница 4: ...activities Avoid use in situations with heavy loads vibrations or impacts Key Text with this symbol contain important safety information WARNING CAUTION Text with this symbol highlights potential sou...

Страница 5: ...he Case 1 2 3 4 5 6 7 1 bebionic Radio Frequency RF Dongle 2 bebionic3 Hand 3 bebionic Tool Kit 4 bebionic in car Charger 5 bebalance Programming Software 6 bebionic3 User Guide 7 bebionic3 Technical...

Страница 6: ...o Hand Base 4 7 8 125mm 4 3 4 121mm C Max Chassis Width no glove 3 5 8 92mm 3 1 4 84mm D Diameter of Chassis at Wrist 2 50mm 2 50mm Palm Circumference no glove 8 5 8 220mm 8 204mm Maximum Opening Widt...

Страница 7: ...1lb 5oz 591g bebionic3 Medium Hand with EQD Wrist BBHMD QD MF 6 1 151mm 25mm 1lb 8oz 691g bebionic3 Medium Hand with Multi Flex Wrist BBHMD SW 4 6 8 1 2 121mm 12mm 1lb 4oz 570g bebionic3 Medium Hand w...

Страница 8: ...08 2 0 bebionic grips...

Страница 9: ...in their day to day life The hand has two selectable thumb positions opposed and non opposed Opposed places the thumb in opposition to the ngers on the hand allowing grips like Pinch and Power Non Op...

Страница 10: ...use A Program Switch is provided on the back of the hand This has four functions 1 Switch the hand ON and OFF A single press for approximately three seconds will switch the hand OFF A single press for...

Страница 11: ...and counterclockwise to move the thumb away from the palm The effect will not be observed while making the adjustment and can only be observed by resetting the hand i e switching the hand OFF and ON a...

Страница 12: ...ly contacts the index nger See section 2 3 2 4 Opposed Thumb is opposite to the ngers Tripod When the thumb is opposed the hand closes into Tripod Grip with index and middle fingers meeting the thumb...

Страница 13: ...les are also held easily and securely Active Index With the thumb opposed Active Index Grip will grasp the handle of an object with the middle ring and little fingers and secure the grip with the thum...

Страница 14: ...capture or reposition of the object being gripped This pattern is ideal for carrying paper or letters using a spoon and for holding a thin flat object such as a plate a credit card or a key Finger Po...

Страница 15: ...tons and switches Column is also the recommended grip for dressing as the thumb is kept out of the way Mouse The thumb and little nger close to hold the side of the mouse with the middle and ring nger...

Страница 16: ...r the thumb to be manually repositioned by the practitioner technician so that the thumb only contacts the index nger Precision Open With the thumb opposed the index nger meets the static thumb allowi...

Страница 17: ...rs from the relaxed hand position Finger Adduction The ngers of the bebionic hand move together naturally as the ngers close This allows the user to securely grip thin objects such as cutlery or a too...

Страница 18: ...ay be fully opened to provide a at palm suitable for carrying a tray or a plate Relaxed Hand The thumb is set to the non opposed position and partially driven in toward the palm All the ngers are driv...

Страница 19: ...19 be yourself...

Страница 20: ...20 3 0 bebionic glove...

Страница 21: ...eed to drive the hand and increases battery life The standard glove is supplied in 19 skin tones and in a solid black color bebionic gloves provide an additional barrier to dirt dust and moisture ingr...

Страница 22: ...OFF by pressing down the program switch on the back of the hand for approximately three seconds Move the thumb into the opposed posi tion 5 Fold up the base of the glove 6 Pull the glove over the han...

Страница 23: ...for four seconds to exit glove donning mode and begin using the hand 3 4 Press and hold the program switch for four seconds The hand will automatically drive into glove mode The thumb will be in the p...

Страница 24: ...nning mode by pressing the program switch for 4 seconds Roll up the base of the glove and pull the whole glove off from the back of the hand Do not pull the ngers individually as this can damage the g...

Страница 25: ...emove the nails bonded with this adhesive will permanently damage the glove Cleaning General soiling can be removed with soap and water After cleaning remove all traces of the cleaner and wipe the sur...

Страница 26: ...26 4 0 connections and compatibility...

Страница 27: ...owing section of this document However if you have a question regarding a speci c build not outlined in this document please contact customer services The bebionic hand may also be used with other man...

Страница 28: ...ith a rotating action The EQD wrist allows the patient to quickly rotate and remove or attach terminal devices as required Low pro le connector for applications where there is a long residual limb A s...

Страница 29: ...rist Offers passive wrist movement in all directions and the ability to lock in 30 exion 30 extension or a neutral position Lateral deviation remains available while the wrist is locked in the preferr...

Страница 30: ...1300S Typical Capacity 1300mAh Voltage Nominal 7 4V Size 8 6mm x 3mm x 66 5mm x2 3 4 x 1 8 x 2 5 8 BBI 2200S Typical Capacity 2200mAh Voltage Nominal 7 4V Size 18 5mm x 18 5mm x 70mm x2 3 4 x 3 4 x 2...

Страница 31: ...dule is tted to the battery and requires an additional connector cable to link to the hand The connector cable for EQD hands as shown on page 32 is supplied with each battery For short wrist or fricti...

Страница 32: ...tailed below is suitable for a bebionic hand with an Electronic Quick Disconnect Wrist Including Multi Flex Wrist BBI 1300S bebionic 7 4V 1300mAh split cell battery BBI 2200 bebionic 7 4V 2200mAh batt...

Страница 33: ...A72 will need to be ordered separately for hands with short wrist Components not to scale The system detailed below is suitable for a bebionic hand with a Short Wrist BBI 2200S bebionic 7 4v 2200mAh s...

Страница 34: ...d with all alternative electrodes wrists and control systems Any damage caused will be held to be outside the design parameters of our system The statements above do not imply that RSLSteeper SteeperU...

Страница 35: ...Motion Control Utah Arm 3 and 3 Supplier Part Number Name Otto Bock 10S17 13E205 Wrist Rotator and MyoRotronic Motion Control MC Wrist Rotator Wrist Mode Type Compatibility Program 1 Two Electrodes Fa...

Страница 36: ...36 5 0 operating instructions...

Страница 37: ...customize the function of the hand Language selection electrode con guration with a simulation preview and simpli ed menu systems are all included in bebalance3 Note Installation disc is af xed to the...

Страница 38: ...features choose between one or two electrode or transducer control and enable the Auto Grip feature All are able to enhance the users natural control capability bebalance software also provides the a...

Страница 39: ...tained it is advisable to back up the previous version before proceeding The bebalance application requires Net Framework 4 0 to operate If not already installed it can be installed from the CD by ope...

Страница 40: ...rget just insert the dongle and return to the entry screen menu or go to the con guration screen and click on Radio When the RF Dongle has been found the message box will show RF Dongle Found highligh...

Страница 41: ...ay will CLOSE the hand a con gurable delay of 50ms to 2 0s will allow repeat CLOSES if required To switch between default grip and alternative grip apply a further OPEN signal within variable time up...

Страница 42: ...o allow the user to try them Successful alterations can be loaded and saved in the hand s memory Details of individual iteration once saved are with the patient s ID in the History le Hand Con guratio...

Страница 43: ...ed position These grips are activated by pressing the program switch on the back of the hand As with the primary grip patterns the user can switch between the default and the alternate grip by applyin...

Страница 44: ...Horizontal Buttons Reset If at any time you want to start again with your con guration clicking Reset reverts all the controls to the default settings for the current mode Save The hand can be con gur...

Страница 45: ...rom the hand and stored in the history le Following a read you are required to enter a patient identi er This enables retrieval of the speci c patient con guration if desired Radio This enables connec...

Страница 46: ...mum levels can be adjusted using the relevant sliders The strength of the signal is displayed as a green vertical bar with the peak signal indicated as single green line On closing the form the settin...

Страница 47: ...l remove the previous trace and automatically start a new one If you wish to stop the trace at any point click Stop The display period can be changed by selecting the appropriate time period 10 20 or...

Страница 48: ...1 The Upper control sets the time that the signal has to exist to be identi ed as a CLOSE signal The Lower control sets the value for the period a signal has to be held once considered an OPEN signal...

Страница 49: ...49 be con dent...

Страница 50: ...50 additional information warranty returns and repairs 6 0...

Страница 51: ...nstructions are set out in this section For further advice on any repairs please contact your local supplier or email bebionic rslsteeper com For both parts and service please mention that the hand is...

Страница 52: ...f the hand Tilt the hand backward and stretch the strap over the bottom of the EQD 5 6 Pull the gaiter away from the palm of the hand making sure that the thumb cable stays in position Taking care to...

Страница 53: ...ained within the toolkit loosen the screws in the rear cover Note Do not completely remove Slide gaiter aps out from under the rear cover 7 8 Completely remove the gaiter from the hand Note the thumb...

Страница 54: ...umb Pay special attention to the thumb cable position Ensure the thumb section of the gaiter is pushed all the way to the gap shown in the image above Carefully push the ap all the way home un der the...

Страница 55: ...over the EQD Line up the sections of the hand highlighted with the sections of the gaiter when pushing the gaiter aps under the rear cover Pay special attention to the thumb cable position Retighten...

Страница 56: ...pivot pin between the clevis and leadscrew with the bebionic pivot removal pliers Part 508 27430 Alternatively the pivot can be removed with the pin punch provided in the tool kit Keeping the nger exe...

Страница 57: ...slot in the proximal Ensure that the curve of the clevis link curves away from the nger towards the palm Bend nger forward to align the holes for the pivot pin Ensure that the nger exes fully forward...

Страница 58: ...compliance with the standard is monitored by a programme of internal and external audits All individual products are marked indicating that they comply with the requirements of the Medical Devices Di...

Страница 59: ...xposure compliance This transmitter must not be co located or operating in conjunction with any other antenna or transmitter Changes or modi cations not expressly approved by the party responsible for...

Страница 60: ...ith the maximum permissible gain and required antenna impedance for each antenna type indicated Antenna types not included in this list having a gain greater than the maximum gain indicated for that t...

Страница 61: ...dule In that case the nal end product must be labeled in a visible area with the following Contains Transmitter Module IC 10634A RSLRFBB or Contains IC 10634A RSLRFBB The OEM of the AMB2520 RSL Module...

Страница 62: ...etteur par Industrie Canada Dans le but de r duire les risques de brouillage radio lectrique l intention des autres utilisateurs il faut choisir le type d antenne et son gain de sorte que la puissance...

Страница 63: ...be found underneath Glove Policy Cosmetic gloves are only replaceable under warranty where the failure is due to a manufacturing fault as we have no control over the environment in which they are used...

Страница 64: ...64 Notes...

Страница 65: ...65 Notes...

Страница 66: ...66 Notes...

Страница 67: ...67 Installation Disc Please refer to page 39 for installtion instructions...

Страница 68: ...nit 7 Hunslet Trading Estate Severn Road Leeds LS10 1BL UK Tel 44 0 870 240 4133 e enquiries rslsteeper com 3619 Paesanos Parkway Suite 200 San Antonio TX 78231 USA Tel 210 481 4126 e inquiries steepe...

Отзывы: