TC305/TC303 - Service
B - 5
lost. After reassembling, refer to chapter
"Special sensor definition"
in
part C of the Instruction manual to fix your additional sensor to a special
sensor number.
3.2 Updating the software version
TC305/TC303 software is stored in two EPROM memories, one on each
processor board on the bottom of the calibrator. The left side (front view)
processor board is named 'main processor board' and the right side processor
board is 'temperature processor board'. The EPROMs are labeled 'Main
Processor' and 'Temp. Processor' respectively. To install a new software version,
follow the updating instructions delivered with the new EPROM(s).
3.3 Installing software options
To be able to install software options, you must purchase the license for the
option from us. The license is delivered in form of security codes. To be able to
generate the codes, we need the serial number of the calibrator as well as the
serial number of the analog i/o assembly of the calibrator.
To read the serial numbers, select the configuration menu entry
CONF
<OTHERS>
(quick select
q
CONF
q
STO
)
To the prompt
C.CODE?
enter the code
101 ENT
For example the display
C.330
E.10234
means that the serial number of the calibrator is 330 (should be the same as
marked on the connection panel) and the serial number of the analog i/o
assembly is 10234.
Write the numbers on paper and press
q
ESC
to exit. Contact us for further
operations.