TC305/TC303 - Calibration
C - 11
7.1 Calibration with thermocouple
Use a good quality thermocouple of type E, J, K or T. We recommend the type E
because of its high output voltage. Do not use extension wires or plug. Connect
the thermocouple to the TC305/TC303 thermocouple connector block. Place the
other end of the sensor into the calibration bath (usually an icebath) at least one
hour earlier to assure good temperature stabilization. Allow the temperatures in
the thermocouple connector block to settle at least five minutes after connection
before starting the calibration.
Select the required thermocouple type with the up arrow key (key SENSOR[7] is
numeric in this mode). The upper display shows now the measured calibration
bath temperature in default temperature units.
The lower display prompts
REF?
Key in the correct temperature, wait until the measurement has been stable over
25 seconds and press
q
ENT
. If you know the error of the thermocouple in this
measurement, you should take it into account.
Example
: The measured temperature is 0
°
C and the reference junction
temperature is 26
°
C. From the calibration of this thermocouple you know, that in
this situation it gives the same voltage as a standard thermocouple gives in
0.08
°
C. In this case you should enter 0.08
°
C instead of 0
°
C.
If there were not at least 21 successive readings within certain limits before you
pressed
q
ENT
, TC305/TC303 flashes
NOT
STABLE
In this case, make sure that the reading is stable and press
q
ENT
again.