42 - Active Terminals and Connections
Note:
To ensure good contact between the device under test
and the test leads, we recommend using the alligator clips
provided with MC6-T.
If you get "+OVER" or “-OVER” error messages, check the
connections. If necessary, use 2-wire resistance measurement
to check the wiring.
Resistance Measurement
For R1 and R3 terminals:
The two leftmost terminals are used in 2-wire systems. MC6-T
automatically checks the connection and displays the wiring system found
(2-wire, 3-wire or 4-wire).
Note:
For the 3-wire system measurements use the connector
marked with "3W".
R2
terminal always uses 4-wire measurement. See
of the female connector in MC6-T
Figure 24: Resistance measurement terminals. Range -1 … 4040 Ohm