16 - MC6-T Safety
Warnings Concerning Electrical
Measurement and Generation
Warning:
MC6-T's OUT section terminals are protected against
over voltage and over current as far as it has been possible
without affecting the accuracy. Do not however connect signals
that are above the measurement range of the selected Function.
Maximum output voltage from MC6-T's terminals is below 30
V. If you, however, connect together voltages from the IN and
OUT sections or if you connect external voltages to MC6-T, the
resulting voltage may be high enough to be hazardous.
There is no galvanic isolation between the USB, Ethernet and
External Pressure Module connectors.
However there is a galvanic isolation between MC6-T's IN and
OUT sections. R3 connector is also galvanically isolated from
the other connectors. It is for functional purposes only. Do not
exceed 60 VDC / 30 VAC / 100 mA between any terminals.
Functional isolation is not intended for protection against
transient overvoltages. Do not connect measuring terminals to
circuits where transient overvoltages are possible.
Warnings Concerning Pressure
General Warnings Concerning Pressure
Measurement
Warning:
The maximum pressures of optional pressure T-
hoses may not be exceeded.
Applying higher pressure than the maximum pressure can be
hazardous.
We recommend the use of the optional pressure hose sets.
When using other hoses and connectors, make sure they are
high quality products that withstand the used pressure.