82
Advanced Operation and Configuration
Data Logging
General
The Data Logging utility is a firmware option that allows you to col-
lect data using MC5P. The collected data may then be viewed, trans-
ferred to a personal computer (PC) and printed using a utility shipped
together with MC5P if the Data Logging option is purchased.
The maximum amount of measuring channels can be up to seven
(valid only if your MC5P includes a sufficient amount of measuring
modules/ports).
Data Logging may be configured to collect data using different trig-
gers and saving methods as described further on.
Configuring
To start configuring the Data
Logging (starting from Basic
Mode), press
D
/
Menu
,
C
/
Others
,
5
/
Data Logging
.
The upper part of the Data Logging
configuration window is reserved for
seven measurement channels and
the lower part for general settings.
The following subchapters describe
how they are configured.
Assigning Measurement Ports to Channels
Move the cursor to one of the meas-
urement channel rows in the Data
Logging configuration window.
Press
C
/
Edit
to open the following
channel settings window: Select the
measurement port first. The other
settings depend on the port setting.
To clear a channel, select port
None
in the channel settings window. To
remove all channel settings, select
D
/
MENU
,
1
/
Clear All Channels
in
the Data Logging configuration win-
dow.
10.11.2003 8:12
DATA LOGGING SETUP
1
Trigger
None
2
None
3
None
4
None
5
None
6
None
7
None
Save Method
Interval [s]
Samples
Time [h min s]
Periodic
Instant Value
2
10
0
Mode
Basic
Logging
Data
Edit
MENU
0
20
10.11.2003 8:13
DATA LOGGING SETUP
Channel
Port
Pressure Type
Unit
1
Abs
bar
P3:B
Cancel
OK
Edit