Editing Instrument Data
MC4 User Guide
171
FIELD
(cont.)
DESCRIPTION
Use Separate
Switch Reset Point
Error Limits
Sometimes the set point of a switch is
more important than the reset point (or vice
versa). Then you may want to set a sepa-
rate reject error limit for the reset point. To
do this, set this check box.
This field is visible for switches only
Reject if >
Active only if previous Check box is set.
Enter a reject error limit for the reset point.