Display Mode and Special Measurements
MC4 User Guide
89
Difference
In Difference measurement, two similar measurements (e.g.
pressure measurement using both the internal and an external
pressure module) are subtracted.
The Difference replaces the main
measurement value. The main
measurement (minuend, indicated
with a circled A) as well as the
secondary measurement value
(subtrahend, indicated with a
circled B) are shown on the
additional info row as seen in the
picture.
Difference measurement is active,
text "
Difference
" is shown above
the Difference value.
S e t u p
S e t u p
M
e n u
N o n e
2 1 . 0 9 . 2 0 0 4 1 1 : 1 9 : 2 2
b a r
D i f f e r e n c e
- 0 . 0 0 3
0 . 4 4 8
0 . 4 5 1
+
A
-
B
Notes.
If anything else was selected for display on the additional info row,
the main measurement of the window replaces the previous
additional info data.
Also: if you select other items to be displayed on the additional info
row while Difference display is active, the items replace the
Difference related data.
Beware of the problems that may result
in not seeing the true measurement values
.
Difference measurement can be activated only if two similar
measurements are available, and neither of them is used for
any other measurement.
Warnings.
Keep in mind that when the displayed Difference reading is
small compared to the actual reading, a significant part of the
Difference value may be measurement error. See the
specifications
for
measurement
errors
at
the
actual
measurement level.
If the measuring spans of the selected ports are different, make
sure you do not exceed the measurement range of either port.
Содержание MC4
Страница 46: ...Service 34 MC4 User Guide Empty ...
Страница 106: ...Tools Menu 94 MC4 User Guide Empty ...
Страница 134: ...Related Information 122 MC4 User Guide Empty ...
Страница 195: ...Appendixes Appendix 1 Technical Data 184 Appendix 3 Index 203 ...
Страница 223: ...Notes MC4 User Guide 211 Notes ...
Страница 224: ...Notes 212 MC4 User Guide ...