2.   Removal of Old Panel

2.1  Remove Ohmeda Front Panel

1.  Prior to removal and installation of control 

panels, verify zone valve is shut o

  and line 

to control panel is not pressurized.

2.  Loosen set screws of on-o

  valve handle and 

remove handle.

3.  Remove regulator adjusting screw and knob 

by unscrewing from regulator.

4. Unscrew lens covers from gauges and 

remove.

5.  Remove screws from cover plate. Remove 

cover plate. Discard

6.  Disconnect both tubing connections at top of 

box.

7.  Remove screws in mounting brackets and 

remove control panel manifold from box. 

Discard.

Figure 1

CAUTION:

Prior to removal and installation of control 

panels, verify zone valve is shut o

 and line to 

control panel is not pressurized.

Cover Plate

Regulator 
Adjusting Screw 
and Knob

Tubing

Connections

On-Off Valve Handle

Gauge Lens

 Covers

Regulator

Mounting

Bracket

3.  Installing New Panel

3.1 Installing 

Retro

fi

 t Kit

1. Remove 

fl

 are nut assembly (supplied with 

retro

fi

 t  kit) from polyethylene bag and install 

to 

fi

 ttings at top of rough-in box.

2. Remove protective dust caps from hose 

ends. Insert hose assembly end into left inlet 

connector inside mounting box.  Push hose 

adapter through collet into connector until it 

bottoms out.  Pull back on the hose assembly 

to verify proper attachment.

Figure 2

Figure 3 

Figure 4

Wall Mount Gas Control Panel

4107 9027 47.00

2

Содержание 4107 9027 47.00

Страница 1: ...Installation Operation and Maintenance Instructions Ohmeda Retrofit Gas Control Panel Part number 4107 9027 47 00 March 16 2020...

Страница 2: ......

Страница 3: ...l number Option s included Any information service or spare parts requests should include the unit model number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Phone 803 817 5600 Fa...

Страница 4: ......

Страница 5: ...ts 1 1 4 Environmental Declarations 1 1 5 About Gas Control Panels 1 2 Removal of Old Panel 2 2 1 Remove Ohmeda Front Panel 2 3 Installing New Panel 2 3 1 Installing Retrofit Kit 2 4 Testing 3 4 1 Tes...

Страница 6: ......

Страница 7: ...ered as well as the environmental perils and dangers to health during the recycling and disposal of the unavoidable rates of non recyclable materials BeaconMed s products for the most part consist of...

Страница 8: ...e 1 CAUTION Prior to removal and installation of control panels verify zone valve is shut o and line to control panel is not pressurized Cover Plate Regulator Adjusting Screw and Knob Tubing Connectio...

Страница 9: ...Operating Instructions 1 Connect surgical tool supply hose to outlet connection on control panel or remote outlet served by control panel 2 Adjust surgical tool operating pressure turning pressure ad...

Страница 10: ...Carefully pull front panel with control assembly out of mounting box Support control assembly in position which allows full view of all connections 3 Adjust control pressure to 160 psig Leak test all...

Страница 11: ...5 4107 9027 47 00 Wall Mount Gas Control Panels Notes...

Страница 12: ...1059 Paragon Way Rock Hill SC 29730 888 4 MEDGAS 888 463 3427 Fax 803 817 5750 www beaconmedaes com...

Отзывы: