3.  Using the same installation procedure from 

step 2, connect other hose end to front 

panel’s inlet connector.

4.  If remote outlet is used, remove pipe plug 

from outlet tee 

fi

 tting and install provided 

1/4” MNPT x 3/8” OD tube swivel elbow into 

outlet tee 

fi

 tting.

Figure 6

4. Testing

4.1 Testing 

Instructions

1.  Open the zone valve that supplies gas to the 

control panel.

1.  Turn pressure adjustment knob clockwise 

until reading of 160 psig is shown on the 

outlet pressure gage.  Pressure will increase 

correspondingly with adjustment.

2.  Listen at panel for sound of escaping gas 

which would indicate leaks.

3.  Turn pressure adjustment knob completely 

counterclockwise until o

 .  Gas venting 

inside the panel shall be heard.  Gas is being 

vented through the self-relieving pressure 

control regulator. Observe outlet pressure 

gage, it shall drop to zero and 

fl

 ow of gas 

shall terminate.

4.2 Operating 

Instructions

1.  Connect surgical tool supply hose to outlet 

connection on control panel or remote outlet 

served by control panel. 

2. Adjust surgical tool operating pressure 

turning pressure adjustment knob clockwise 

to increase or counterclockwise to decrease 

pressure, as indicated on outlet pressure 

gage.  For proper operating pressure, consult 

tool manufacturer’s recommendations.

NOTE:
It is best to adjust pressure while gas is 

fl

owing 

through tool.

5.  When remote outlet is used, insert hose 

assembly end into right remote outlet 

connector inside mounting box.  Push hose 

adapter through collet into connector until 

it bottoms out.  Pull back on hose assembly 

to verify proper attachment.  Using same 

installation procedure connect other hose 

end to front panel’s remote outlet connector.

6. Using leak detection solution, leak test 

threaded connection of 

fl

 are nut and 

fi

 tting 

in rough-in box.

7.   Position front panel over rough-in box and 

screw into place using (4) #6-32 x 1-1/2” 

screws provided with retro

fi

 t kit. 

Figure 5

Wall Mount Gas Control Panels

4107 9027 47.00

3

Содержание 4107 9027 47.00

Страница 1: ...Installation Operation and Maintenance Instructions Ohmeda Retrofit Gas Control Panel Part number 4107 9027 47 00 March 16 2020...

Страница 2: ......

Страница 3: ...l number Option s included Any information service or spare parts requests should include the unit model number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Phone 803 817 5600 Fa...

Страница 4: ......

Страница 5: ...ts 1 1 4 Environmental Declarations 1 1 5 About Gas Control Panels 1 2 Removal of Old Panel 2 2 1 Remove Ohmeda Front Panel 2 3 Installing New Panel 2 3 1 Installing Retrofit Kit 2 4 Testing 3 4 1 Tes...

Страница 6: ......

Страница 7: ...ered as well as the environmental perils and dangers to health during the recycling and disposal of the unavoidable rates of non recyclable materials BeaconMed s products for the most part consist of...

Страница 8: ...e 1 CAUTION Prior to removal and installation of control panels verify zone valve is shut o and line to control panel is not pressurized Cover Plate Regulator Adjusting Screw and Knob Tubing Connectio...

Страница 9: ...Operating Instructions 1 Connect surgical tool supply hose to outlet connection on control panel or remote outlet served by control panel 2 Adjust surgical tool operating pressure turning pressure ad...

Страница 10: ...Carefully pull front panel with control assembly out of mounting box Support control assembly in position which allows full view of all connections 3 Adjust control pressure to 160 psig Leak test all...

Страница 11: ...5 4107 9027 47 00 Wall Mount Gas Control Panels Notes...

Страница 12: ...1059 Paragon Way Rock Hill SC 29730 888 4 MEDGAS 888 463 3427 Fax 803 817 5750 www beaconmedaes com...

Отзывы: