Do not place any objects on top of the units
Penetrating liquids may result in a fire or electric shock.
Connection
No contact to a patient must occur when handling the cables.
Do not hurt yourself, when moving the display
The display can be tilted backwards and forwards. Please, pay attention
not to hurt yourself, when moving the display. Fingers or small objects
may get stuck at the bottom of the display.
When moving the display up and down (height adjustment), make sure you
do not squeeze your hand or any other object. The minimum distance
between the display edge and the bottom is only 59 mm.
Caution
Incorrect installation may result in extensive damage to property.
Installation should be carried out by trained personnel
When installing your medical electrical system with our products in
an environment with patients, please observe the safety requirements
of EN 60601-1 (IEC 60601-1) for "Specifications for the safety of
medical electrical systems" in order to prevent injury to patients and
users of your systems.
Take appropriate measures to particularly ensure that discharge
currents remain below the required limits: Appropriate measures:
- Disconnecting devices for signal input or output unit
- Use of a safety transformer
- Use of additional PE conductor
Only use the signal cables and interface cables specified by the
manufacturer for the installation.
Use power cables with a PE contact. Only insert into sockets with a
PE contact.
For certain applications, the video earth can be separately connected
to the PE via the additional PE connection in the plug panel (observe
IEC 60601-1).
Close the plug panel using the provided cover , and secure using the