11
11
3-WAY IC SYRINGE
113-400
ITEM
1
2
3
4
5
6
7
8
9
10
11
12
PART DESCRIPTION
Syringe Tip ..................................
Syringe Tip Nut............................
O-Ring, 006 .................................
O-Ring, 228-040..........................
Syringe Tip Adapter.....................
O-Ring, 003 .................................
Syringe Head...............................
Syringe Button Repair Kit ............
Pin, Syringe Cartridge .................
Plastic Washer, 8-32....................
Barb, 8-32 x 1/16’’ .......................
Syringe Handle
...........................
PART NUMBER
113-030
113-402
017-006
017-023
113-227
017-003
113-405
113-403
113-008
003-030
022-040
113-236
QTY
1
1
2
1
1
1
1
2
1
2
2
1
Quick-disconnect Syringe Tip Assembly 113-230
1. Remove the existing hex nut. A hex nut wrench has been
provided.
2. Remove the syringe tip adapter using the hex wrench.
3. Insert the new syringe tip adapter into the syringe head and
tighten using the hex wrench.
4. Place the two (2) O-Rings into the syringe tip nut.
5. Screw the syringe tip nut onto the syringe tip adapter and
tighten using the hex wrench.
6. Push the syringe tip into the syringe tip nut until it snaps into
place.
Replace Syringe Tip
1. Pull the dirty syringe tip out and insert a clean syringe tip into
the syringe tip nut until it snaps into place.