Chapter 4: Maintenance
77
Procedure
1
Choose Cytometer (Instrument) > Cleaning Modes > Clean Flow Cell.
2
If you have a Loader installed, remove the carousel.
3
When prompted, manually install a tube containing approximately 2 mL of
cleaning solution onto the SIT, and click OK.
A progress message is displayed during the cleaning.
4
After the completion message displays, wait 5 minutes.
Waiting allows BD FACSClean solution to dissolve deposits in the flow cell
cuvette.
5
Click OK.
6
Remove the tube from the SIT.
7
Clear BD FACSClean solution from the flow cell and fluidics lines:
•
Before running setup or samples, run Fluidics Startup.
•
To shut down without running more samples, run Fluidics Shutdown.
BD FACSDiva software
BD FACSCanto clinical software
BD FACSCanto clinical software
BD FACSDiva software