BD FACSCanto II Flow Cytometer Reference Manual
Light Scatter
When a cell or particle passes through a focused laser beam, laser light is
scattered in all directions (Figure A-4). Light that scatters roughly in the same
direction as the laser beam is called forward scatter (FSC); light that scatters
roughly perpendicular to the laser beam is called side scatter (SSC). FSC and SSC
intensities indicate physical properties of cells:
•
FSC—indicates relative differences in the size of the cells or particles
•
SSC—indicates relative differences in the internal complexity or granularity
of the cells or particles
Figure A-4
Forward scatter (FSC) and side scatter (SSC)
laser
side scatter
forward scatter