Chapter 3: BD CARV II Touch Pad Controller
47
Figure 3-9
Presets Screen
Calibration
Each CARV II instrument is individually calibrated at the factory before
shipping. Each position of each device is aligned to the light path. Most of the
factory calibration involves internal hardware that is not accessible in the
finished instrument. However, each position on the filter wheels, the position of
the prism in the light path, and the position of the confocal disk in the light path
can be calibrated using the touch pad. Because calibrating the devices within the
CARV II can adversely affect the optical performance, only qualified personnel
should perform these procedures.
Calibrating the Prism
NOTICE
Once the calibration for the prism has been changed, there is no way to
restore the factory setting without optically aligning the instrument; therefore,
use caution when adjusting.
Press and hold the binoc button for a few seconds to activate the calibration
function and adjust the position of the prism in the light path. Once the
calibration function is activated, the Main Screen will appear as in Figure 3-10.