Display
Action
BATTERY LOW
Low battery threshold sensed; remaining
battery run time is limited; if cancelled will
reoccur every ten minutes if not plugged in.
• Connect to power supply.
• Check power cable.
AC POWER FAIL
AC power disconnected or failed.
• Reconnect to power supply.
VTBI DONE
Intended VTBI completed. When
KVO
or
CONTINUE
are selected as the
END RATE
.
• Set new VTBI or clear VTBI.
AIR-IN-LINE
Air detected in infusion set at the start of
infusion. Set not fitted correctly into air in
line detector.
• Ensure set is fitted correctly in the air
in line detector.
• Assess air in infusion set.
• Check fluid level in drip chamber.
• Check fluid level in container.
SET CLOCK
Date/time not set.
• Qualified service personnel must set
date/time.
• Press
CANCEL
softkey to continue.
TITRATION
Rate titration not confirmed.
• Confirm or cancel new rate.
RATE LOCK
Rate lock not confirmed.
• Select
YES
or
NO
as required.
LOG FAILURE
Unable to update event log.
• Qualified service personnel may need
to service the Pump.
SET SERIAL NUMBER
Serial number not set.
• Contact qualified service personnel
to set the serial number.
NEAR END OF INFUSION
The Pump is nearing the end of the
infusion. This value can be configured.
• Prepare next infusion, if required.
KVO
Occurs after a
VTBI DONE
warning is
cancelled if
KVO
was selected as the
END
RATE
setting.
• Set new infusion or stop treatment