ii) Insert Safety Clamp and pressure disc into retainer.
Safety Clamp
iii) Ensure infusion set is fully inserted into tubing guide and all air is removed
from the infusion set.
Tubing Guide
iv) Use minimal stretching of the silicone segment when loading or priming
the infusion set.
7. Close door and open roller clamp. Ensure no drops are falling in the drip chamber. If
drops are falling in the drip chamber then clamp the infusion set immediately, open
door and verify correct loading of the infusion set.
Pushing the Safety Clamp Slider completely into Safety Clamp Frame
may lead to uncontrolled flow to the patient. Therefore, always close
the roller clamp before pushing in the Safety Clamp Slider, if required.
Loading an Inf
usion Set c
ontinued