
15
EN
USER MANUAL
XTA
8. CALIBRATION
Turn on the scale. Keep pressed
key during the self test.
Display will show “
Pn
” briefly PIN is activated. Press the PIN number to get into the parameter menu.
•
The default PIN number is “0000”
•
Press numeric key 0 four times, display will be shown
Pn - - - -
• Press
key to confirm, it will enter into the parameter settings and will to show “
F0 CAL
”.
Then press the calibration switch to access to calibration.
• Press
key to confirm for calibration, display will be show “
UNLOAD
”
•
Remove the goods from the platform and make sure STABLE indication has been enable.
• Press
key to confirm.
•
Display will be show
,CAP LoAdXX.
•
If necessary change the calibration weight value by pressing
key and place the callibration weight on the platform.
• Press
key to confirm when display gets stable, display will be show
PASS.
Finished calibration and display will come to
FO CAL.
•
Remove calibration weigjt from the platform.
• Press
key for to go back to the weighing mode.
9. RS 232 OUTPUT
1. SPECIFICATIONS:
RS-232 output of weighing data
ASCII code
1200~115200 Baud
8 data bits
No Parity
2. RS-232 (9PIN D TYPE CONNECTOR)
Pin 2
RXD
Input
Receiving data
Pin 3
TXD
Output
Transmission data
Pin 5
GND
-
Signal ground