Configuration
Page 66
Auto Syringe AS50 Infusion Pump Operator’s Manual
07-19-B3-430
Idle Alarm
This option enables the pump to issue a continuous audible alarm when a
PUMP IS IDLE
condition occurs. If the option is disabled, the pump beeps 15 times at two minute
intervals when a
PUMP IS IDLE
condition occurs.
Near End Alert (NEAR END)
This option enables the pump to issue an alert during an infusion prior to an expected
VOL LIMIT
or
EMPTY
alarm. The option specifies the number of minutes prior to the alarm
that the expected alert will be issued and the number of beeps in the alert.
The form of a Near End alert display is
xx
MIN VOL LIM
,
xx
MIN EMPTY
or
xx
MIN
(protocol).
The value
xx
is the configured number of minutes. The first and second forms apply to
impending
VOLUME LIMIT
or
EMPTY
conditions when no protocol has been selected. The
third form applies to either condition when a protocol has been selected. Press
CONFIRM
to view the current settings.
Note:
When an infusion is started, no
NEAR END
alert will occur if fewer than the config-
ured number of minutes remain before an
EMPTY
or
VOL LIMIT
alarm is due to
occur. For example, if the
NEAR END
alert is configured for 10 minutes and 9 min-
utes remains before the volume limit is reached, no
NEAR END
alert will occur
when
START
is pressed.
Bolus Complete Alert (BOLUS COMPLETE)
This option enables the pump to issue an alert when a bolus completes.
Dose Complete Alarm (DOSE COMPLETE)
This option enables the pump to issue an alarm when a dose completes in SINGLE
DOSE mode.
Volume Limit Alarm (VOLUME LIMIT)
This option applies to mL/hr and mL/min modes only. During initial programming, the
pump prompts the operator to enter a volume limit value in mL. The Volume Limit feature
triggers an alarm after the pump has delivered the pre-programmed volume of drug. With
this option enabled, a
NEAR END
alert is issued when the Volume Limit will be reached
within a preconfigured time period at the current rate of infusion. See “Near End Alert” on
page 66.
Identifier
The Identifier is not a configuration group, but is included here for reference purposes.
The Identifier is a short message (up to eight characters) that is displayed for three
seconds following the Lamp Test. The Identifier can be created or edited during the pump
configuration process.