Auto Syringe AS50 Infusion Pump Operator’s Manual
Page 63
Configuration
Configurable Options
The option groups and configurable options are listed here and in the table above. Each
option is described in the following paragraphs. The option is displayed exactly as listed in
the paragraph title unless otherwise noted.
Protocol Library Group (
VIEW LIBRARY
)
This option group displays the drug protocols that are enabled. The pump displays a
VIEW
(library name) prompt for each enabled protocol library. A pending
Y
response is
also displayed. To select a library, press
CONFIRM
to accept the pending response. To
skip a library, press
▲
or
▼
to change the
Y
to
N
and press
CONFIRM
.
When a library has been selected, the pump displays a
VIEW
(protocol name) prompt for
each enabled protocol. A pending
Y
response is also displayed. To begin reviewing a
protocol, press
CONFIRM
to accept the pending response. To skip a protocol, press
▲
or
▼
to change the
Y
to
N
and press
CONFIRM
.
For each selected protocol, the default mode (if any) is displayed in the text field (for
example,
MG/KG/HR
), and concentration, dose and rate default values (if any) are shown
in their usual fields.
Press
CONFIRM
to view the default bolus size.
BOLUS SIZE
is displayed in the text field.
Press
CONFIRM
again to view the bolus maximum rate.
BOLUS MAX RATE
is displayed in
the text field.
Note:
Dashes are displayed whenever a default value is not specified by the protocol.
If the protocol specifies bolus prevention,
NO BOLUS
is displayed instead of
BOLUS SIZE
.
If no defaults are specified by the protocol,
ALL MODES
is displayed when the pro-
tocol is initially selected.
Press
CONFIRM
to advance to the next protocol. When the last protocol in a library has
been presented for viewing, the next
VIEW
(library name) prompt is displayed.
See the Auto Syringe AS50 Infusion Pump Technical Manual for instructions on creating
and editing protocols.
Modes Group (
VIEW MODES
)
This option group displays the name of each enabled infusion mode. Each listing shows
only the mode name (for example
ML/HR
) in the text field. For a list of the supported
infusion modes, see the “Configuration Review Table” on page 62.
Syringes Group (
VIEW SYRINGES
)
This option group displays the enabled syringe sizes for each enabled syringe
manufacturer.