SINGLE NEEDLE
| Terminology
In single needle the patient’s blood is removed from, and returned to, the single access point in a
cycle with two phases;
•
The Arterial phase.
The venous blood line is clamped, and blood is withdrawn from the patient
into the arterial blood line. The pressure and volume in the circuit increases.
•
The Venous phase.
The arterial blood line is clamped, and the blood that entered in the arterial
phase is returned to the patient via pressure and gravity.
Single needle parameters;
•
Mean blood flow.
Is the effective mean blood flow rate during the complete cycle.
•
Stroke volume.
Is the blood volume which passes through the dialyzer during a cycle. The
higher the stroke volume the more efficient the treatment.
HC10139 |
Содержание AK 98
Страница 1: ...AK 98 Machine Introduction Training Material ...
Страница 2: ...Introduction Module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapters 2 3 ...
Страница 14: ...14 THE OPERATOR S PANEL Touch Screen Layout HC10139 ...
Страница 31: ...Preparation module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapters 4 12 ...
Страница 49: ...49 Preparing the machine IT connectivity Setting station ID Using patient ID as a label HC10139 ...
Страница 52: ...Treatment module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapters 4 8 ...
Страница 86: ...Troubleshooting module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapter 3 Alarm handbook ...
Страница 100: ...Baxter AK 98 and BiCart are trademarks of Baxter International Inc or its subsidiaries End of troubleshooting module ...
Страница 101: ...Single needle module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapter 5 ...
Страница 110: ...Baxter and AK 98 are trademarks of Baxter International Inc or its subsidiaries End of Single Needle module ...
Страница 111: ...Hygienic maintenance module AK 98 Dialysis Machine Reference AK 98 Operator s Manual 3 xx Chapters 10 11 ...