TROUBLESHOOTING
| Air in Venous Drip Chamber
When the
blood level in the venous chamber drops below the level of the air detector
, then alarm
#100 Air in venous drip chamber is triggered.
•
Check the blood circuit connections are tight.
•
Use the timer or confirm icon in the alarm message to restart the blood pump, at the same time
increase the blood level in the venous chamber using the level adjustment knob.
Recommended blood level