Preventive Maintenance
23
SIGMA Spectrum Infusion System
Pump Operating Software Version 6.02 and 6.05
Preventive Maintenance
For Use With MDL Version 6
6.
Air Detection Test
Pass criteria:
All steps completed successfully.
Pump goes into Air in Line alarm.
(Rate = 100 mL/hr VTBI = 50 mL)
7.
Battery Capacity Test
Perform the Battery Capacity Test to determine the storage capacity of the
battery.
NOTE:
Test time approximately 9 hours.
Start Time Date: ____________
Time “Start to dead” = ______Hrs. ______Min.
End Time Date: _____________
Pass criteria
Standard Battery
■
Used:
≥
6 hours from “Start to dead”
■
New:
≥
8 hours from “Start to dead”
Wireless Battery
■
Used:
≥
3 hours from “Start to dead”
■
New:
≥
4 hours from “Start to dead”
Start Time Date: ____________
Time “Low to dead” = ______Hrs. ______Min.
End Time Date: _____________
Pass criteria
Standard Battery
■
Used:
≥
31 minutes from “Low to dead”
■
New:
≥
31 minutes from “Low to dead”
Wireless Battery
■
Used:
≥
31 minutes from “Low to dead”
■
New:
≥
31 minutes from “Low to dead”
Tests Performed
Pass
Fail