2
Preventive Maintenance
Manual 41394
SIGMA Spectrum Infusion System
Revision A
Preventive Maintenance
When using the Annual Preventive Maintenance Check Sheet:
■
Enter the Hospital/Facility name in the space provided.
■
Enter the Biomed name and date of the test.
■
Perform the inspections, tests, and performance evaluations described in this chapter.
■
Record the test result findings.
■
Return any Pump to Baxter that fails any inspection or test.
Pump Identification
Serial Number and Operating Software Version
The Pump’s serial number and operating software version are located in two areas, the Power
Status screen and the User Options screen.
Record and verify the serial number and software version of the Pump using one of the methods
described in this section.
To verify the Pump serial number and software version from the Power Status screen:
1. Press
ON/OFF
to power on the Pump.
2. View the current software version and serial number.
3. Record the software version and serial number on the Preventive Maintenance Check Sheet.
4. Check the label on the bottom of the Pump. Verify the serial number is the same.
To verify the Pump serial number and software version from the User Options menu:
1. Press
ON/OFF
to power on the Pump.
2. From the Select Care Area screen, press the
options menu
soft key (software version 6.05)
or
Help
then the
Options
soft Key (software version 6.02).
3. Select
User Options
and press
OK
.
4. Select
View Information
and press
OK
.
5. Select
Pump Information
and press
OK
to display the Pump Info screen.
6. Record the software version and serial number on the Preventive Maintenance Check Sheet.
7. Check the label on the bottom of the Pump. Verify the serial number is the same.
8. Press the
exit
soft key four times to return to the Care Area screen.
Network Connectivity
To verify current Network connectivity:
1. From the Select Care Area screen, press the
options menu
soft key (software version 6.05)
or
Help
then the
Options
soft Key (software version 6.02).